Answers are grounded in this paper and cite their sources.
BACKGROUND: Conventional endoscopes have the disadvantages of causing pain to patients because of their large dimensions and allowing incomplete inspection because of their limited power. METHODS: An active inchworm-like endoscopic robot has been designed, which is based on electromagnetic localization and synthesis of magnetic field strength vectors. RESULTS: The efficiency of wireless power transmission can be kept above 0.9% when the power-receiving coil is at arbitrary position and orientation. The position error of localization is < 10 mm and the average error of orientation angle is 2 degrees. When the robot moves through part of the porcine colon, images are transmitted outwards continuously at a speed of 30 fps and the temperature rise of the colon is < 1 degrees C. By simulation, the electromagnetic exposures to the human body are lower than international standards. CONCLUSIONS: The endoscopic robot is a promising tool for use by doctors, although further experiments and improvements need to be done.
No takes yet. Share an insight, caveat, or question.
Li et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: