Answers are grounded in this paper and cite their sources.
Diabetes is a serious disease with a huge number of patients worldwide. Glucose levels in people with diabetes are above 6.6 mM in blood samples and 200 μM in saliva samples. Metal-organic frameworks (MOFs) have outstanding properties for glucose sensors, such as a large surface area and being rich in active sites. In this research, a silver [email protected] ([email protected]) composite was synthesized through one-pot synthesis using a microwave at 130 °C for 1 hour at 200 W. The results showed that the 4%[email protected] carbon paste electrode obtained better sensor performance than Ni-BTC with a limit of detection (LoD) of 14.73 μM, a sensitivity of 6584.89 μA mM−1 cm−2, and a linear range of 10–1250 μM. The 4%[email protected] carbon paste electrode also had better stability than Ni-BTC and had good reproducibility. Glucose detection tests on salivary samples showed that the 4%[email protected] carbon paste electrode could be used to measure glucose levels in salivary samples.
No takes yet. Share an insight, caveat, or question.
Amri et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: