Answers are grounded in this paper and cite their sources.
Methodological study demonstrates simultaneous fluorimetric quantification of cortisol, corticosterone, and testosterone in plasma, highlighting high analytical specificity and ease of use.
The simultaneous estimation of steroids in plasma was carried out by the assay of cortisol, corticosterone, and testosterone. The method entails separation by means of thin layer chromatography, followed by conversion to a fluorophore and fluorimetric measurement. Its major advantages are its high specificity, its ability to detect unknown substances, and the ease with which it can be performed. The method has acceptable levels of accuracy and precision and the normal values obtained by it compare well with those given by methods in general use.
No takes yet. Share an insight, caveat, or question.
Mattingly et al. (1989) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: