Answers are grounded in this paper and cite their sources.
We report a fast and simple high-speed liquid-chromatographic assay for serum retinol. Only 100 mul of serum is required. The lower detection limit is 50 mug/liter; linearity was demonstrated up to 1.50 mg/liter. On analyzing a serum pool eight times, a CV of 2.5% was obtained. Values by this method are compared with results obtained by a flurometric method [Clin. Chem. 16, 766(1970)].
No takes yet. Share an insight, caveat, or question.
Ruyter et al. (1976) studied this question.