Key result
Continuous noninvasive blood pressure monitoring showed acceptable accuracy for mean blood pressure (bias -3.3 ± 5.6 mm Hg) but poor accuracy for systolic blood pressure versus invasive monitoring.
Why the study?
Error grid analysis was recently proposed to compare blood pressure obtained by two measurement methods, motivating this comparison of continuous noninvasive blood pressure with invasive blood pressure and investigation of confounding risk factors.
Does continuous noninvasive blood pressure monitoring accurately reflect invasive blood pressure in adult patients undergoing general anesthesia?
Observational (n=60)
Does continuous noninvasive blood pressure monitoring accurately reflect invasive blood pressure in adult patients undergoing general anesthesia?
Continuous noninvasive blood pressure monitoring shows larger clinical discrepancies in mean blood pressure compared to systolic blood pressure when evaluated by error grid analysis, with older age and longer cuff inflation time increasing the risk of dangerous discrepancies.
Answers are grounded in this paper and cite their sources.
Continuous noninvasive MBP monitoring may show clinically relevant discrepancies versus invasive readings; leaves open need for prospective validation using error grid analysis.
BACKGROUND: Error grid analysis was recently proposed to compare blood pressure obtained by 2 measurement methods. This study aimed to compare continuous noninvasive blood pressure (CNBP) with invasive blood pressure (IBP) using the error grid analysis and investigate the confounding risk factors attributable to the differences between CNBP and IBP. METHODS: Sixty adult patients undergoing general anesthesia were prospectively enrolled. Simultaneous comparative data regarding CNBP and IBP were collected. The Bland-Altman analysis was conducted to compare CNBP and IBP for systolic blood pressure (SBP) and mean blood pressure (MBP; acceptable accuracy: mean bias <5 mm Hg; standard deviation <8 mm Hg). The clinical relevance of the discrepancies between CNBP and IBP was evaluated by the error grid analysis, which classifies the differences into 5 zones from "no risk" (A) to "dangerous risk" (E). Additionally, an ordinal logistic regression analysis was performed to evaluate the relationship between the risk zones for MBP, classified by the error grid analysis and covariates of interest. RESULTS: A total of 10,663 pairs of CNBP/IBP were finally analyzed. The Bland-Altman analysis showed an acceptable accuracy with a bias of -3.3 ± 5.6 mm Hg for MBP but a poor accuracy with a bias of 5.4 ± 10.5 mm Hg for SBP. The error grid analysis showed the proportions of zones A to E as 96.7%, 3.2%, 0.1%, 0%, and 0% for SBP, respectively, and 72.0%, 27.9%, 0.1%, 0%, and 0% for MBP, respectively. The finger cuff missed 23.9% of epochs when SBP <90 mm Hg and 55.3% of epochs when MBP <65 mm Hg. The ordinal logistic regression analysis revealed that older age (adjusted odds ratio for decade: 1.54, 95% confidence interval [CI], 1.15-2.08; P = .004) and length of time from the initiation of finger cuff inflation (adjusted odds ratio for 60 minutes: 1.40, 95% CI, 1.13-1.73; P = .002) were significant factors of being in a more dangerous zone of the error grid. CONCLUSIONS: The error grid analysis revealed the larger clinical discrepancy between CNBP and IBP in MBP compared with that in SBP. Old age and longer finger cuff inflation time were significant factors of being in a more dangerous zone of the error grid, which could affect the hemodynamic management during surgery.
No takes yet. Share an insight, caveat, or question.
A 2022 study conducted an observational in General anesthesia (n=60). Continuous noninvasive blood pressure (CNBP) monitoring vs. Invasive blood pressure (IBP) monitoring was evaluated on Discrepancy between CNBP and IBP evaluated by Bland-Altman and error grid analysis. Continuous noninvasive blood pressure monitoring showed acceptable accuracy for mean blood pressure (bias -3.3 ± 5.6 mm Hg) but poor accuracy for systolic blood pressure versus invasive monitoring.