Key result
Cuffless beat-to-beat monitoring overestimates invasive brachial systolic BP by ~34 mmHg.
Why the study?
An accurate continuous cuffless BP monitor would provide an alternative to invasive continuous and 24-h intermittent cuff-based BP monitors, requiring accuracy comparison against both invasive and brachial cuff BP.
Does a cuffless beat-to-beat blood pressure monitor (SOMNOtouch NIBP) accurately reflect invasive brachial artery blood pressure and oscillometric brachial cuff blood pressure in patients undergoing coronary angiography or percutaneous coronary intervention?
Observational (n=14)
No
Does a cuffless beat-to-beat blood pressure monitor (SOMNOtouch NIBP) accurately reflect invasive brachial artery blood pressure and oscillometric brachial cuff blood pressure in patients undergoing coronary angiography or percutaneous coronary intervention?
The SOMNOtouch NIBP cuffless beat-to-beat blood pressure monitor substantially overestimates systolic and diastolic blood pressure compared to invasive brachial artery and standard cuff measurements during coronary angiography/PCI and is unsuitable for clinical or research use.
Answers are grounded in this paper and cite their sources.
Clinicians should avoid this cuffless device for BP monitoring; cross-sectional data leaves open its accuracy in other settings and populations.
BACKGROUND: The availability of an accurate continuous cuffless blood pressure (BP) monitor would provide an alternative to both invasive continuous BP and 24-h intermittent cuff-based BP monitors. We investigated the accuracy of a cuffless beat to beat (BtB) device compared to both invasive BP (iBP) and brachial cuff BP (cBP) measurements. METHODS: Patients undergoing clinically indicated coronary angiography (CA) and/or percutaneous coronary intervention (PCI) were recruited. After calibration to an initial cBP reading, BP was measured simultaneously using a BtB device (SOMNOtouch NIBP), brachial artery iBP, and cBP at two time points. RESULTS: The study was terminated early due to a significant bias. Recordings from 14 participants (11 males, mean age 68.4 years) were analysed. Readings from BtB BP were higher than iBP. The bias between BtB BP and iBP was 34.3 mmHg (95%CI: 27.0, 41.5) and 23.6 mmHg (95%CI: 16.8, 30.4) for SBP and DBP respectively. A similar bias was seen between BtB BP and cBP, but cBP and iBP were largely in agreement. CONCLUSIONS: In patients undergoing CA/PCI, significant differences were detected between BtB BP and both invasively measured and cuff BP. The non-invasive BtB BP measurement device tested is not suitable for clinical or research use.
No takes yet. Share an insight, caveat, or question.
Moharram et al. (2020) conducted an observational in Patients undergoing clinically indicated coronary angiography and/or percutaneous coronary intervention (n=14). Cuffless beat-to-beat blood pressure measurement (SOMNOtouch NIBP) vs. Invasive brachial artery blood pressure (iBP) and oscillometric brachial cuff blood pressure (cBP) was evaluated on Bias (mean difference) in systolic blood pressure (SBP) between cuffless beat-to-beat BP and invasive brachial artery BP (MD 34.3 mmHg, 95% CI 27.0, 41.5). Cuffless beat-to-beat blood pressure monitoring (SOMNOtouch NIBP) overestimated invasive brachial systolic BP by 34.3 mmHg and diastolic BP by 23.6 mmHg.