Why the study?
Validation of cuffless blood pressure monitoring technologies against the highest physiological reference standard, invasive arterial line monitoring, remains limited and has not been systematically synthesised.
Do cuffless blood pressure monitoring devices provide accurate systolic and diastolic blood pressure measurements compared with invasive arterial line monitoring in patients in intensive care unit or immediate post-operative settings?
Do cuffless blood pressure monitoring devices provide accurate systolic and diastolic blood pressure measurements compared with invasive arterial line monitoring in patients in intensive care unit or immediate post-operative settings?
Cuffless blood pressure devices demonstrated minimal mean bias for both systolic and diastolic blood pressure compared with invasive arterial line monitoring in acute care environments, supporting the feasibility of noninvasive continuous haemodynamic assessment in the ICU.
Supports cuffless BP monitoring in ICU; extends validation of noninvasive devices versus arterial lines.
Background Cuffless blood pressure (BP) monitoring technologies are increasingly proposed for continuous and remote cardiovascular assessment. However, validation against the highest physiological reference standard, invasive arterial line (ART) monitoring, remains limited and has not been systematically synthesised. Purpose To perform a pooled analysis of studies evaluating cuffless BP devices against invasive ART measurements in acute care settings. Methods As part of an ongoing systematic review of cuffless BP technologies, we performed a predefined subgroup analysis of studies directly comparing cuffless devices with invasive ART monitoring. Eligible studies were conducted in intensive care unit (ICU) or immediate post-operative settings and reported within-patient mean differences (MD) with dispersion measures. Where paired standard deviations were not explicitly reported, they were derived from Bland–Altman limits of agreement, and corresponding standard errors were calculated as SD/√n. Random-effects meta-analyses were performed separately for systolic BP (SBP) and diastolic BP (DBP) using restricted maximum likelihood (REML) with Hartung–Knapp adjustment. Risk of bias was assessed using the QUADAS-2 tool. The systematic review was registered in PROSPERO (CRD420261326740). Results Four studies (n = 164) conducted in ICU or immediate post-operative settings with continuous monitoring (≈4–25 h) met inclusion criteria. For DBP, the pooled mean difference was 0.39 mmHg (95% CI 0.09–0.69) with no heterogeneity (I² = 0%). For SBP, the pooled mean difference was 1.42 mmHg (95% CI −3.44 to 6.28) with moderate heterogeneity (I² = 64.9%). Leave-one-out analyses confirmed robust DBP findings and identified a single outlier as the primary driver of SBP heterogeneity, with its exclusion eliminating heterogeneity and attenuating the pooled effect. Overall risk of bias was low to moderate. Conclusion In pooled analysis of invasive arterial line validation studies, cuffless BP devices demonstrated minimal mean bias for both SBP and DBP in acute care environments. These findings represent synthesis within the highest physiological reference context currently available and support the feasibility of cuffless monitoring in high-acuity settings. Given the risks and resource burden associated with invasive arterial lines, accurate noninvasive continuous monitoring may offer a safer and more scalable alternative for haemodynamic assessment in the ICU. Ongoing analyses will evaluate performance across broader ambulatory and outpatient populations.DBP cufflessSBP cuffless