Key result
Cuffless pulse wave BP estimators lack sufficient validity for clinical use despite person-specific calibration.
Why the study?
Does cuffless blood pressure monitoring using pulse wave travel times accurately estimate systolic and diastolic blood pressure compared to cuff-based measurements in healthy adults?
Population
41 young, healthy participants, mean age 21.1 years, 66% female, based in the Netherlands. Exclusion…
Comparison
Cuffless ambulatory blood pressure monitoring… vs Ambulatory cuff-based blood pressure monitoring…
Design
Cross-sectional
Authors
Loading...
Cuffless PWTT BP estimators should not enter clinical use; leaves open need for prospective validation beyond person-specific calibration.
Observational (n=41)
No
Does cuffless blood pressure monitoring using pulse wave travel times accurately estimate systolic and diastolic blood pressure compared to cuff-based measurements in healthy adults?
Effect estimate: MAE 10.1-12.9 mmHg
Cuffless blood pressure monitoring using pulse wave travel times shows high reliability but insufficient validity for clinical or stress research use, even with person-specific calibration.
Schoenmakers et al. (2026) conducted an observational in Healthy (n=41). Cuffless ambulatory blood pressure monitoring (pulse wave travel times) vs. Cuff-based blood pressure monitoring was evaluated on Convergent validity (Mean Absolute Error) and estimation accuracy within 10 mmHg of reference values (MAE 10.1-12.9 mmHg). Cuffless pulse wave travel time-based blood pressure estimators showed strong test-retest reliability (ICC > 0.7) but insufficient convergent validity (MAE 10.1-12.9 mmHg) for clinical use, even with person-specific calibration.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: