Key result
A combination of epitope-distinct anti-cTnI monoclonal antibodies recognizing the N-terminal half of cTnI yielded reliable detection and greater sensitivity for cTnI in AMI patients.
Combining epitope-distinct monoclonal antibodies against the N-terminal half of cTnI improves the sensitivity of enzyme immunoassays for detecting acute myocardial infarction.
May enhance cTnI immunoassay sensitivity for AMI; leaves open clinical translation from animal models.
Human cardiac troponin I (cTnI) is one of the most specific biomarkers for detection of acute myocardial infarction (AMI). To formulate immunoassay kits for rapid immunodiagnosis of AMI, monoclonal antibodies with high affinity and specificity were generated against cTnI and subsequently tested through a series of experiments. C57BL/6 mice were immunized with cTnI as the immunogen and cell fusions with myeloma cells of BALB/c origin were performed to generate hybridomas. The supernatants of the hybridoma cell culture were routinely screened for antibody secretions against intact cTnI and synthetic peptides from the N-terminal half of cTnI (amino acid residues N1-30, N24-40, N59-79, and N80-95). Monoclonal antibodies specific to different epitope regions were then determined and selected, according to their respective affinity and specificity, for formulation of enzyme immunoassay kits. The results of this study found that most of the selected antibodies revealed comparable binding affinity to cTnI and to the corresponding synthetic peptides. Optimal sandwich enzyme immunoassays with high sensitivity could be achieved through proper combinations of the epitope-distinct monoclonal antibodies in different capture-detection pairs; signal enhancements were frequently observed when a mixture of epitope-distinct anti-cTnI monoclonal antibodies was used for coating. This indicates that a combination of epitope-distinct anti-cTnI monoclonal antibodies recognizing the N-terminal half of cTnI yield reliable detection and greater sensitivity for cTnI in AMI patients.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2015) studied Acute myocardial infarction. Monoclonal antibodies against human cardiac troponin I was evaluated on Binding affinity and specificity to cTnI and synthetic peptides. A combination of epitope-distinct anti-cTnI monoclonal antibodies recognizing the N-terminal half of cTnI yielded reliable detection and greater sensitivity for cTnI in AMI patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: