Key result
An explicit finite element simulation method predicted guidewire-induced vascular deformations during EVAR with an average error of 4.6 ± 2.5 mm compared to intraoperative measurements in 9 patients.
A finite element simulation method can predict vascular deformations during EVAR guidewire insertion with an average error of 4.6 mm compared to intraoperative measurements.
Supports simulation feasibility for EVAR planning; leaves open clinical use pending larger validation.
Deformations of the vascular structure due to the insertion of tools during endovascular treatment of aneurysms of the abdominal aorta, unless properly anticipated during the preoperative planning phase, may be the source of intraoperative or postoperative complications. We propose here an explicit finite element simulation method which enables one to predict such deformations. This method is based on a mechanical model of the vascular structure which takes into account the nonlinear behavior of the arterial wall, the prestressing effect induced by the blood pressure and the mechanical support of the surrounding organs and structures. An analysis of the model sensitivity to the parameters used to represent this environment is done. This allows determining the parameters that have the largest influence on the quality of the prediction and also provides realistic values for each of them as no experimental data are available in the literature. Moreover, for the first time, the results are compared with 3D intraoperative data. This is done for a patient-specific case with a complex anatomy in order to assess the feasibility of the method. Finally, the predictive capability of the simulation is evaluated on a group of nine patients. The error between the final simulated and intraoperatively measured tool positions was 2.1 mm after the calibration phase on one patient. It results in a 4.6 ± 2.5 mm in average error for the blind evaluation on nine patients.
No takes yet. Share an insight, caveat, or question.
Gindre et al. (2015) studied Aneurysms of the abdominal aorta (n=9). Explicit finite element simulation method vs. 3D intraoperative data was evaluated on Error between final simulated and intraoperatively measured tool positions. An explicit finite element simulation method predicted guidewire-induced vascular deformations during EVAR with an average error of 4.6 ± 2.5 mm compared to intraoperative measurements in 9 patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: