Key result
The designed Analog-to-Information Converter based on Compressed Sensing successfully reconstructed real electrocardiograms and electromyograms with compression rates of 8 and 10, respectively.
A novel Analog-to-Information Converter based on Compressed Sensing enables highly efficient acquisition and reconstruction of biomedical signals like ECGs with high compression rates and low power consumption.
We report the design and implementation of an Analog-to-Information Converter (AIC) based on Compressed Sensing (CS). The system is realized in a CMOS 180 nm technology and targets the acquisition of bio-signals with Nyquist frequency up to 100 kHz. To maximize performance and reduce hardware complexity, we co-design hardware together with acquisition and reconstruction algorithms. The resulting AIC outperforms previously proposed solutions mainly thanks to two key features. First, we adopt a novel method to deal with saturations in the computation of CS measurements. This allows no loss in performance even when 60% of measurements saturate. Second, the system is able to adapt itself to the energy distribution of the input by exploiting the so-called rakeness to maximize the amount of information contained in the measurements. With this approach, the 16 measurement channels integrated into a single device are expected to allow the acquisition and the correct reconstruction of most biomedical signals. As a case study, measurements on real electrocardiograms (ECGs) and electromyograms (EMGs) show signals that these can be reconstructed without any noticeable degradation with a compression rate, respectively, of 8 and 10.
No takes yet. Share an insight, caveat, or question.
Pareschi et al. (2015) studied this question. Analog-to-Information Converter based on Compressed Sensing was evaluated on Signal reconstruction SNR and compression rate. The designed Analog-to-Information Converter based on Compressed Sensing successfully reconstructed real electrocardiograms and electromyograms with compression rates of 8 and 10, respectively.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: