Does a prototype intravascular ultrasonic imaging catheter provide accurate imaging of human arteries compared to MR imaging and tissue specimens?
Does a prototype intravascular ultrasonic imaging catheter provide accurate imaging of human arteries compared to MR imaging and tissue specimens?
This early in vitro study demonstrates the feasibility and potential diagnostic value of intravascular ultrasound imaging for evaluating human arteries.
Answers are grounded in this paper and cite their sources.
Supports preclinical feasibility of intravascular ultrasound; leaves open human validation before clinical consideration.
A small-aperture, prototype ultrasonic imaging system, typical of the size necessary to be embedded in a catheter, was developed to evaluate the feasibility of intravascular ultrasonic imaging catheters (UICs). The evaluation included in vitro imaging of postmortem samples of human femoral and iliac arteries with the UIC prototype and a high-resolution magnetic resonance (MR) proton imager. Excellent correlations between results from the UIC images, MR images, and tissue specimens were demonstrated. Although the current prototype is too large to be useful in examination of smaller vessels such as coronary arteries, the signal-to-noise ratio and resolution obtained indicate that imaging with intravascular UICs is feasible and may have significant diagnostic value.