Lead contamination in drinking water is a primary concern in public health, but it is difficult to monitor by end-users. Here, we provide a rapid and power-free microfluidic particle dam which enables visual quantification of lead ions (Pb 2+ ) by the naked eye. GR-5 DNAzyme with extended termini can connect magnetic microparticles (MMPs) and polystyrene microparticles (PMPs) by DNA hybridization, forming “MMPs-GR-5-PMPs”. When Pb 2+ is present, GR-5 is cleaved, resulting in an increasing number of free PMPs. To visually count the free PMPs, the solution is loaded to a capillary-driven microfluidic device that consists of a magnetic separator to remove the MMPs-GR-5-PMPs, followed by a particle dam that traps and accumulates the free PMPs into a visual bar with growing length proportional to the concentration of lead. The device achieved a limit of detection at 2.12 nM (0.44 ppb), high selectivity (>20,000-fold) against other metal ions, high tolerance to different pH and water hardness, and is compatible with tap water with a high recovery rate, enabling visual quantification and user-friendly interface for rapid screening of water safety.