Key result
PDMAA-grafted PU showed the lowest extent of platelet deposition among the tested polymeric materials in a rabbit ex vivo shunt model, with no trace of aggregate detachment.
Why the study?
Does surface modification of polymeric tubes reduce platelet deposition in a rabbit ex vivo shunt model?
Does surface modification of polymeric tubes reduce platelet deposition in a rabbit ex vivo shunt model?
Answers are grounded in this paper and cite their sources.
PDMAA-grafted PU surfaces significantly reduce platelet deposition compared to standard polymeric materials in an ex vivo shunt model, suggesting potential for heparin-free short-term blood-contacting devices.
In an attempt to develop blood-contacting tubes that can be applied for short-term uses with a reduced heparin concentration or, ideally, without heparinization, we evaluated the blood compatibility of polymeric materials with a rabbit ex vivo shunt model. The shunt tubes employed were made of silicone, plasticized poly(vinyl chloride) (PVC), and segmented poly(ether urethane) (PU). In addition, two kinds of surface-modified tube were used: poly(vinyl alcohol) (PVA)-coated PVC and poly(dimethylacrylamide) (PDMAA)-grafted PU. The ex vivo shunt results correlated well with protein adsorption and platelet adhesion in vitro. The following order for the extent of platelet deposition was given, irrespective of the blood-contacting duration: PDMAA-grafted PU < PVA-coated PVC < PU < silicone, PVC. It is likely that many platelet aggregates detached from the PVA-coated PVC surface. For PDMAA-grafted PU, no trace of detachment of aggregates could be detected on any of the SEM photographs. The number and morphology of blood cells adhered onto the tube surfaces during ex vivo shunting were dependent on the kind of polymer surfaces, the blood exposure time, and the flow rate of blood.
No takes yet. Share an insight, caveat, or question.
Fujimoto et al. (1993) studied Blood compatibility of polymeric materials. Polymeric materials (silicone, PVC, PU, PVA-coated PVC, PDMAA-grafted PU) vs. Compared against each other was evaluated on Platelet deposition and blood compatibility. PDMAA-grafted PU showed the lowest extent of platelet deposition among the tested polymeric materials in a rabbit ex vivo shunt model, with no trace of aggregate detachment.