Answers are grounded in this paper and cite their sources.
Plasma process control applications require acquisition of diagnostic data at a rate faster than the characteristic timescale of perturbations to the plasma. Diagnostics based on optical emission spectroscopy of intense emission lines permit rapid noninvasive measurements with low-resolution (∼1 nm), fiber-coupled spectrographs, which are included on many plasma process tools for semiconductor processing. Here the authors report on rapid analysis of Ar emissions with such a system to obtain electron temperatures, electron densities, and metastable densities in argon and argon/mixed-gas (Ar/N2, Ar/O2, Ar/H2) inductively coupled plasmas. Accuracy of the results (compared to measurements made by Langmuir probe and white-light absorption spectroscopy) are typically better than ±15% with a time resolution of 0.1 s, which is more than sufficient to capture the transient behavior of many processes, limited only by the time response of the spectrograph used.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: