Why the study?
Does Doppler control compared to optical adjustment optimize the flow pattern of the Abiomed BVS 5000 assist device in an in vitro model?
Does Doppler control compared to optical adjustment optimize the flow pattern of the Abiomed BVS 5000 assist device in an in vitro model?
In an in vitro model of the Abiomed BVS 5000, Doppler flow control at the inflow site is necessary to guarantee maximal stable flow and prevent high atrial pressures and retrograde flow associated with optical adjustment.
Doppler inflow control may stabilize hemodynamics with the Abiomed BVS 5000; leaves open translation to clinical practice.
The Abiomed BVS 5000 is an automatic volume-driven paracorporeal pulsatile assist device providing left, right or biventricular support. The paracorporeal position allows optical adjustment of filling volumes of the device, which determines the output of the system. A procedure to adjust for maximal stable flow has not yet been established. In vitro measurements have been performed to assess the flow and pressure characteristics of the Abiomed BVS 5000 by raising the preload in 5 mmHg steps before running the system. Doppler probes were placed at the inflow and outflow lines of the pump. After setting the afterload at 80 mmHg the assist device was started. Two measurements were performed to find optimal flow (based on Doppler control and optical adjustment). (1) By Doppler control a stable flow pattern was found at a preload of 25 mmHg with a mean atrial pressure of 5 mmHg and a mean flow of 5.3 +/- 0.7 l/min (mean +/- standard deviation) at the inflow and outflow sites (the console flow was 4.8 +/- 0.4 l/min with a frequency of 61.8 +/- 2.0 l/min). (2) Optical adjustment of the pump height gave rise to a preload of 35 mmHg where we recorded a maximal atrial pressure of 107 +/- 5.8 mmHg, a maximal retrograde flow of -4.3 +/- 1.2 l/min at the inflow and -1.2 +/- 0.4 l/min at the outflow site. The mean flow at the inflow and outflow sites was 5.1 +/- 0.5 l/min (the console flow was 4.6 +/- 0.3 l/min with a frequency of 59.6 +/- 2.6 Hz). At an initial afterload of 60 and 40 mmHg the system showed the same qualitative behaviour, but the results were less accurate. Optical adjustment of the pump height may result in an atrioventricular valve insufficiency with undetected retrograde flow and high atrial pressures. We conclude that a Doppler flow probe must be placed at the inflow site to guarantee maximal stable flow.
No takes yet. Share an insight, caveat, or question.
Lachat et al. (1999) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: