Key result
Cosine motion profile, higher ejection fraction, higher pulsatile rate, and counterpulsation decreased platelet deposition but increased hemolysis and platelet activation in VAD simulations.
Why the study?
Do different operating conditions influence blood damage in a pulsatile ventricular assist device?
Do different operating conditions influence blood damage in a pulsatile ventricular assist device?
Different operating conditions of pulsatile VADs have varying effects on blood damage indices, suggesting that optimizing these parameters could reduce blood damage in clinical applications.
Because of pulsatile blood flow's benefit for myocardial recovery, perfusion of coronary arteries and end organs, pulsatile ventricular assist devices (VADs) are still widely used as paracorporeal mechanical circulatory support devices in clinical applications, especially in pediatric heart failure patients. However, severe blood damage limits the VAD's service period. Besides optimizing the VAD geometry to reduce blood damage, the blood damage may also be decreased by changing the operating conditions. In this article, a pulsatile VAD was used to investigate the influence of operating conditions on its blood damage, including hemolysis, platelet activation, and platelet deposition. Three motion profiles of pusher plate (sine, cosine, and polynomial), three stroke volumes (ejection fractions) (56 ml [70%], 42 ml [52.5%], and 28 ml [35%]), three pulsatile rates (75, 100, and 150 bpm), and two assist modes (copulsation and counterpulsation) were implemented respectively in VAD fluid-structure interaction simulations to calculate blood damage. The blood damage indices indicate that cosine motion profile, higher ejection fraction, higher pulsatile rate, and counterpulsation can decrease platelet deposition whereas increase hemolysis and platelet activation, and vice versa. The results suggest that different operating conditions have different effects on pulsatile VAD's blood damage and may be beneficial to choose suitable operating condition to reduce blood damage in clinical applications.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2015) studied this question. Different operating conditions (motion profiles, stroke volumes, pulsatile rates, assist modes) was evaluated on Blood damage (hemolysis, platelet activation, and platelet deposition). Cosine motion profile, higher ejection fraction, higher pulsatile rate, and counterpulsation decreased platelet deposition but increased hemolysis and platelet activation in VAD simulations.