Answers are grounded in this paper and cite their sources.
We present a method for measuring ascorbic acid in methanol/trichloroacetic acid extracts prepared from human plasma after enzymatic oxidation of ascorbic acid to dehydroascorbic acid by ascorbate oxidase. Samples were assayed by spectrophotometrically monitoring the kinetics of the concentration-dependent absorbance changes of dehydroascorbic acid with phosphate-citrate-methanol buffers. Ascorbic acid was determined as the difference between dehydroascorbic acid and total ascorbic acid content. The detection limit was < 0.5 mumol/L. The calibration curve was linear (r > 0.995) over the range 0-1000 mumol/L. Analytical recovery of ascorbic acid added to plasma was 93-105%. The between-day variance was < 7%. Comparison of the spectrophotometric determination (y) with a chromatographic procedure (x) gave y = 1.02x - 0.653 (Sylx = 3.61) over the range of physiologically relevant concentrations. Total analysis time is < 10 min per sample and allows the simultaneous analysis of multiple samples.
No takes yet. Share an insight, caveat, or question.
Moeslinger et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: