Answers are grounded in this paper and cite their sources.
OBJECTIVES/HYPOTHESIS: To determine the incidence of cochlear implant failure and to examine surgical and audiological outcomes. STUDY DESIGN: Retrospective review, case series. METHODS: This study sought indications for revision surgery, surgical findings, and outcomes, and audiological outcomes in pediatric cochlear implant patients. Pre- and postcochlear reimplantation word recognition performance was analyzed using a modified version of the Pediatric Ranked Order Speech Perception (PROSPER) score. RESULTS: Over a 20-year period, a total of 868 cochlear implants were performed in 578 patients. The overall institutional reimplant rate was 5.9%. The indications for explantation were hard failure (30), soft failure (23), and medical/surgical indication (13). A significant portion of devices belonged to vendor recalled batches (15) or were damaged by head trauma (eight). Full electrode insertion was achieved in all 62 reimplantations. Post-reimplantation Boston Children's Hospital modified PROSER scores were either stable or improved compared to pre-reimplantation scores. CONCLUSIONS: The need for cochlear implant revision/reimplantation is infrequent, but the rate is not inconsequential. Hard and soft device failures account for the majority of reimplants. Surgical complications during reimplantation is low, and post-reimplantation audiological performance is excellent. LEVEL OF EVIDENCE: 4. Laryngoscope, 2619-2624, 2018.
No takes yet. Share an insight, caveat, or question.
Yeung et al. (2018) studied this question.