Key result
The tubular cardiovascular phantom system successfully simulated pulse transit time, with optical sensors showing a sensitivity of -0.3902 ms/mmHg to pulse pressure changes.
Why the study?
Most PTT measurements are performed on human subjects, complicating sensor validation due to BP variability.
A tubular cardiovascular phantom system can successfully simulate pulse transit time variations with pulse rate and pressure, offering a platform for validating cuffless blood pressure sensors.
This paper presents on the development of a tubular cardiovascular phantom system to simulate pulse transit time (PTT). The PTT defined as the delay time between two pulses in one cardiac cycle has been shown to be promising method for cuffless continuous blood pressure (BP) measurement. However most of the PTT measurement was performed on human subjects, thus giving a difficulty in validating sensor performance due to variability of BP. Therefore, a cardiovascular phantom system was proposed for simulate the PTT measurement. An electronic controlled module was developed to control pump operation for pulse generation. Plastic optical fibre (POF) sensors were used to measure the pulse signal on the flexible tube and the results were compared with an in-line pressure sensor. In this experiment, the delay time between two pulses were calculated offline using Matlab software and correlated with pulse pressure. The result demonstrate that the pulse delay time recorded by both sensors decreased with increase of pulse rate and pulse pressure. These results on the phantom study showed similar pattern to the human model, thus indicating that the system is able to simulate PTT for sensor validation purposes.
No takes yet. Share an insight, caveat, or question.
A 2019 study studied Cardiovascular simulation (Phantom study). Tubular cardiovascular phantom system with plastic optical fibre (POF) sensors vs. In-line pressure sensors was evaluated on Pulse transit time (PTT) sensitivity to pulse pressure (ms/mmHg). The tubular cardiovascular phantom system successfully simulated pulse transit time, with optical sensors showing a sensitivity of -0.3902 ms/mmHg to pulse pressure changes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: