Key result
WATCHMAN FLX transfabric leaks occur in up to ~52% of patients without clear predictors.
Why the study?
Contrast flow across a left atrial appendage occlusion (LAAO) device on computed tomography, known as a transfabric leak (TFL), may indicate incomplete device endothelialization, but data on the rate, predictors, and clinical events associated with TFL are limited.
What are the rate, predictors, and clinical events associated with transfabric leaks (TFL) on cardiac CT angiography after left atrial appendage occlusion with the WATCHMAN FLX device?
Cohort (n=56)
No
What are the rate, predictors, and clinical events associated with transfabric leaks (TFL) on cardiac CT angiography after left atrial appendage occlusion with the WATCHMAN FLX device?
Transfabric leaks are detected in 27% to 52% of WATCHMAN FLX patients on 4-month cardiac CT angiography depending on the radiographic definition used, indicating that complete device endothelialization in humans is frequently incomplete at 4 months unlike in preclinical animal models.
Answers are grounded in this paper and cite their sources.
Frequent transfabric leaks without predictors after WATCHMAN FLX; leaves open their clinical impact and need for prospective validation.
Background: Cardiac computed tomography imaging with contrast is being used increasingly to image left atrial appendage occlusion (LAAO) devices. Contrast flow across a device, also known as a transfabric leak (TFL), may indicate a lack of complete LAAO-device endothelialization. The data on the rate, predictors, and clinical events associated with TFL are limited. Methods: All patients who underwent an LAAO-device implantation with a WATCHMAN FLX device and received a postimplantation computed tomography scan were included in this single-centre retrospective cohort study. Patients were classified as either having or not having a TFL, according to 3 currently proposed definitions of TFL. Clinical and procedural differences between the 2 groups were determined. An exploratory univariate logistic regression model to evaluate predictors of TFL was constructed. Results: A total of 56 individuals were included in the cohort. The rate of TFL varied from 27% to 52%, depending on the radiographic definition employed. No clinically important patient or procedural characteristics were noted between the groups with vs without TFL. No predictors of TFL were identified. Six deaths and one stroke occurred during a median follow-up period of 673 days. Conclusions: TFLs occur commonly post-LAAO procedures, suggesting that complete endothelialization of LAAO devices in humans may not be similar to that reported in animal models. Additional study into the best imaging approach to identify TFLs, and clinical events associated with TFLs, is necessary to clarify the significance of TFLs.
No takes yet. Share an insight, caveat, or question.
Bauer et al. (2024) conducted a cohort in Nonvalvular atrial fibrillation undergoing percutaneous left atrial appendage occlusion with WATCHMAN FLX (n=56). Transfabric leak (TFL) on post-implantation cardiac computed tomography angiography after WATCHMAN FLX LAAO vs. Absence of transfabric leak (complete LAAO-device endothelialization) was evaluated on Rate of transfabric leak (TFL) on cardiac computed tomography angiography according to 3 radiographic definitions (criterion 1: LAA HU > 100 and LAA/LA HU ratio > 0.25). Transfabric leaks after WATCHMAN FLX left atrial appendage occlusion occurred in 27% to 52% of patients on 4-month cardiac CT angiography, with no identified clinical or procedural predictors.