Key result
High-energy radiotherapy simulation causes software malfunctions in ~52% of ICDs via scattered neutrons.
Why the study?
Does scattered neutron radiation from high-energy (15 MV) radiotherapy directed far from the thorax cause device malfunctions in explanted pacemakers and implantable cardioverter-defibrillators?
Does scattered neutron radiation from high-energy (15 MV) radiotherapy directed far from the thorax cause device malfunctions in explanted pacemakers and implantable cardioverter-defibrillators?
This experimental phantom study demonstrates that high-energy (15 MV) sub-diaphragmatic radiotherapy produces scattered thermal neutrons capable of causing severe software malfunctions in over half of ICDs and nearly a fifth of pacemakers even without direct beam exposure.
Answers are grounded in this paper and cite their sources.
Warrants pre/post high-energy radiotherapy device checks; leaves open in vivo confirmation and mitigation strategies.
AIMS: Malfunctions of cardiac implantable electronical devices (CIED) have been described after high-energy radiation therapy even in the absence of direct exposure to ionizing radiation, due to diffusion of neutrons (n) causing soft errors in inner circuits. The purpose of the study was to analyse the effect of scattered radiation on different types and models of CIED and the possible sources of malfunctions. METHODS AND RESULTS: Fifty-nine explanted CIED were placed on an anthropomorphous phantom of tissue-equivalent material, and a high-energy photon (15 MV) radiotherapy course (total dose = 70 Gy) for prostate treatment was performed. All devices were interrogated before and after radiation. Radiation dose, the electromagnetic field, and neutron fluence at the CIED site were measured. Thirty-four pacemakers (PM) and 25 implantable cardioverter-defibrillators (ICD) were analysed. No malfunctions were detected before radiation. After radiation a software malfunction was evident in 13 (52%) ICD and 6 (18%) PM; no significant electromagnetic field or photon radiations were detected in the thoracic region. Neutron capture was demonstrated by the presence of the (198)Au((197)Au + n) or (192)Ir((191)Ir + n) isotope activation; it was significantly greater in ICD than in PM and non-significantly greater in damaged devices. A greater effect in St Jude PM (2/2 damaged), Boston (9/11), and St Jude ICD (3/6) and in older ICD models was observed; the year of production was not relevant in PM. CONCLUSION: High-energy radiation can cause different malfunctions on CIED, particularly ICD, even without direct exposure to ionizing radiation due to scattered radiation of neutrons produced by the linear accelerator.
No takes yet. Share an insight, caveat, or question.
Zecchin et al. (2015) studied Cardiac implantable electronic devices (CIED) exposed to scattered radiation (n=59). High-energy photon (15 MV) radiotherapy course simulating prostate cancer treatment vs. Baseline interrogation before radiation (and comparison between ICDs and pacemakers) was evaluated on Device software malfunction after radiation. High-energy (15 MV) radiotherapy simulation caused software malfunctions in 52% (13/25) of implantable cardioverter-defibrillators and 18% (6/34) of pacemakers due to scattered neutrons.