Key result
Radiotherapy photon energy ≥10 MV is linked to ~6.7-fold higher odds of CIED malfunction.
Why the study?
The incidence and clinical predictors of cardiac implantable electronic device malfunction during radiotherapy treatment needed further investigation.
What is the incidence of cardiac implantable electronic device (CIED) malfunction during radiotherapy, and do factors such as photon beam energy ≥10 MV, irradiation above the diaphragm, and device type predict malfunction?
Cohort (n=441)
No
What is the incidence of cardiac implantable electronic device (CIED) malfunction during radiotherapy, and do factors such as photon beam energy ≥10 MV, irradiation above the diaphragm, and device type predict malfunction?
In patients with CIEDs undergoing radiotherapy, device malfunction is driven by photon beam energy ≥10 MV, thoracic/supradiaphragmatic irradiation, and ICD device type rather than cumulative radiation dose, though malfunctions rarely cause clinical symptoms.
Answers are grounded in this paper and cite their sources.
Supports caution with high-energy photons in CIED radiotherapy; leaves open prospective validation of predictors and outcomes.
AIMS: To investigate the incidence of cardiac implantable electronic devices (CIED) malfunction with radiotherapy (RT) treatment and assess predictors of CIED malfunction. METHODS: A 6-year retrospective analysis of patients who underwent RT with CIED identified through the radiation oncology electronic database. Clinical, RT (cumulative dose, dose per fraction, beam energy, beam energy dose, and anatomical area of RT) and CIED details (type, manufacturer, and device malfunction) were collected from electronic medical records. RESULTS: We identified 441 patients with CIED who underwent RT. CIED encountered a permanent pacemaker (PPM) (78%), cardiac resynchronization therapy-pacing (CRT-P) (2%), an implantable cardioverter defibrillator (ICD) (10%), and a CRT-defibrillator (CRT-D) (10%). The mean cumulative dose of RT was 36 gray (Gy) (IQR 1.8-80 Gy) and the most common beam energy used was photon ≥10 megavolt (MV) with a median dose of 7 MV (IQR 5-18 MV). We further identified 17 patients who had CIED malfunction with RT. This group had a higher cumulative RT dose of 42.5 Gy (20-77 Gy) and a photon dose of 14 MV (12-18 MV). None of the malfunctions resulted in clinical symptoms. Using logistic regression, the predictors of CIED malfunction were photon beam energy use ≥10 MV (OR 5.73; 95% CI, 1.58-10.76), anatomical location of RT above the diaphragm (OR 5.2, 95% CI, 1.82-15.2), and having a CIED from the ICD group (OR 4.6, 95% CI, 0.75-10.2). CONCLUSION: Clinicians should be aware of predictors of CIED malfunction with RT to ensure the safety of patients.
No takes yet. Share an insight, caveat, or question.
Azraai et al. (2022) conducted a cohort in Cancer patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy (n=441). Radiotherapy with photon beam energy ≥10 MV vs. Radiotherapy with photon beam energy <10 MV was evaluated on Cardiac implantable electronic device (CIED) malfunction during or after radiotherapy (OR 6.73, 95% CI 1.58-10.76, p=0.007). Photon beam energy ≥10 MV during radiotherapy increased the odds of CIED malfunction, though none of the 17 malfunctions caused clinical symptoms.