Key result
An in vitro bicuspid aortic valve model demonstrated highly unsteady fluid shear stresses on the leaflets compared to a trileaflet model, potentially contributing to accelerated calcification.
The bicuspid aortic valve geometry leads to unsteady, high-frequency shear stresses under physiological flow conditions, which may contribute to the accelerated calcification observed in these patients.
Answers are grounded in this paper and cite their sources.
May implicate unsteady shear in BAV calcification; leaves open clinical translation pending prospective validation.
The bicuspid aortic valve (BAV) is a common congenital malformation of the aortic valve (AV) affecting 1% to 2% of the population. The BAV is predisposed to early degenerative calcification of valve leaflets, and BAV patients constitute 50% of AV stenosis patients. Although evidence shows that genetic defects can play a role in calcification of the BAV leaflets, we hypothesize that drastic changes in the mechanical environment of the BAV elicit pathological responses from the valve and might be concurrently responsible for early calcification. An in vitro model of the BAV was constructed by surgically manipulating a native trileaflet porcine AV. The BAV valve model and a trileaflet AV (TAV) model were tested in an in vitro pulsatile flow loop mimicking physiological hemodynamics. Laser Doppler velocimetry was used to make measurements of fluid shear stresses on the leaflet of the valve models using previously established methodologies. Furthermore, particle image velocimetry was used to visualize the flow fields downstream of the valves and in the sinuses. In the BAV model, flow near the leaflets and fluid shear stresses on the leaflets were much more unsteady than for the TAV model, most likely due to the moderate stenosis in the BAV and the skewed forward flow jet that collided with the aorta wall. This additional unsteadiness occurred during mid- to late-systole and was composed of cycle-to-cycle magnitude variability as well as high-frequency fluctuations about the mean shear stress. It has been demonstrated that the BAV geometry can lead to unsteady shear stresses under physiological flow and pressure conditions. Such altered shear stresses could play a role in accelerated calcification in BAVs.
No takes yet. Share an insight, caveat, or question.
Yap et al. (2012) studied Bicuspid aortic valve. Bicuspid aortic valve (BAV) model vs. Trileaflet aortic valve (TAV) model was evaluated on Fluid shear stresses on the leaflet and flow fields downstream. An in vitro bicuspid aortic valve model demonstrated highly unsteady fluid shear stresses on the leaflets compared to a trileaflet model, potentially contributing to accelerated calcification.
Synapse has enriched 10 related papers. Consider them for comparative context: