Key result
Transcatheter closure of perimembranous VSD using the Amplatzer Duct Occluder II device resulted in complete heart block in 2 cases, which resolved after early surgical removal.
Case Report (n=2)
Complete heart block can occur following transcatheter closure of perimembranous VSD using the Amplatzer Duct Occluder II device, which may be reversed by early surgical removal.
Answers are grounded in this paper and cite their sources.
Raises caution for conduction block with ADO-II PmVSD closure; hypothesis-generating for device safety and reversibility.
Long term follow up studies have shown that transcatheter closure of perimembranous Ventricular Septal Defect (PmVSD) can be done safely and successfully. One of the most serious complications of VSD device closure is complete atrioventricular block which has been reported in 3-18% is various studies following VSD device closure. Amplatzar Duct Occluder II (ADO-II) device is now commonly being used for closure of small to moderate sized PmVSD. Complete heart block has not been reported with ADO-II device. We are reporting 2 cases of complete heart block following transcatheter closure of PmVSD using ADO-II device where sinus rhythm was established following early surgical removal of the device.
No takes yet. Share an insight, caveat, or question.
Ghosh et al. (2017) conducted a case report in Perimembranous Ventricular Septal Defect (PmVSD) (n=2). Transcatheter closure using Amplatzer Duct Occluder II (ADO-II) device was evaluated on Complete heart block. Transcatheter closure of perimembranous VSD using the Amplatzer Duct Occluder II device resulted in complete heart block in 2 cases, which resolved after early surgical removal.
Synapse has enriched 10 related papers. Consider them for comparative context: