Key result
Actual pacemaker longevity was 491+/-92 days less than calculated longevity based on battery cell characteristics and programmed parameters.
Why the study?
Does actual pacemaker longevity match calculated longevity based on battery cell characteristics and programmed parameters in patients undergoing pacemaker replacement?
Observational (n=124)
Does actual pacemaker longevity match calculated longevity based on battery cell characteristics and programmed parameters in patients undergoing pacemaker replacement?
Actual pacemaker longevity is significantly shorter than calculated longevity, suggesting higher than expected current drain for ancillary functions or pre-implantation drain.
Answers are grounded in this paper and cite their sources.
Shorter-than-calculated pacemaker longevity may prompt earlier replacement planning; leaves open prospective validation of battery models and excess drain mechanisms.
BACKGROUND: Although pacemaker manufacturers provide projections on longevity, these projections cannot be relied upon due to the assumptions of output parameters being far in excess of those programmed in clinical practice. OBJECTIVE: The purpose of this review was to compare the actual longevity to the calculated longevity of pacemakers based on battery cell characteristics taking into account individual programmed parameters, mode, degree of usage, and percent pacing. This was also compared to the manufacturers' own projected longevities. METHODS: Patients who had a pacemaker replaced between 1998 and 2003 were included (n = 124). Cell characteristics were obtained from manufacturers and programmed parameters were obtained at each visit. Stepwise calculations were done for each visit to find current drain during each interval, and then were used in a weighted average to find the total average lifetime current drain. This was subsequently used to find a calculated longevity for each pacemaker to be compared to the actual longevity observed. RESULTS: The pacemakers lasted 491+/-92 days (mean+/-SEM) less than calculated. There was also a difference between dual- and single-chamber devices (though not statistically significant). Moreover, it was found that there were significant differences between manufacturers. CONCLUSIONS: There appears to be a significant discrepancy between calculated and actual longevities, confirming that battery depletion occurs earlier than expected. This suggests that current drain expended for ancillary functions may be considerable. Another factor may be pre-implantation drain. Vigilance with programming of outputs, modes, sensors, heart rates, and ancillary functions could potentially extend longevity and postpone/obviate the need for costly repeat surgery with its attended risk of complications. Furthermore, the differences between manufacturers seem to parallel the clinical impressions.
No takes yet. Share an insight, caveat, or question.
Senaratne et al. (2006) conducted an observational in Pacemaker replacement (n=124). Actual pacemaker longevity vs. Calculated pacemaker longevity was evaluated on Difference between actual and calculated pacemaker longevity (MD -491 days). Actual pacemaker longevity was 491+/-92 days less than calculated longevity based on battery cell characteristics and programmed parameters.