Key result
Dual chamber pacemakers demonstrate a median time to battery depletion of ~98 months.
Observational (n=382)
The majority of patients outlive the battery life of their dual chamber pacemakers (median longevity 98.3 months), underscoring the clinical importance of maximizing device longevity.
Answers are grounded in this paper and cite their sources.
Battery longevity data may inform device selection and counseling; leaves open optimal programming strategies for broader utilization.
In 382 patients with three different dual chamber pulse generators, the median time interval to battery depletion was 98.3 months. Cox regression analysis revealed the following variables as significant predictors of battery longevity: programmed pacing rate, energy of the stimulation output, mode of stimulation (i.e., proportion of paced cycles in one or two chambers), battery capacity, and internal sensing current of the pacemaker. Although 27% of all patients died before the service life of the pacemaker was over and despite a rate of premature reoperations of 8.6%, the majority of pacemaker patients (55%) fully used the expected battery life span of the pulse generator. Patients who died before the pacemaker had reached its end of service were significantly older at implantation than patients who survived until pacemaker replacement. The vast majority (92%) of patients received another dual chamber pulse generator when replacement was required. These data underline the need for long-lasting dual chamber devices.
No takes yet. Share an insight, caveat, or question.
Kindermann et al. (2001) conducted an observational in Patients requiring pacemakers (n=382). Dual chamber pulse generators was evaluated on Time interval to battery depletion. Dual chamber pacemakers demonstrated a median time to battery depletion of 98.3 months, with 55% of patients fully utilizing the expected battery life span.
Synapse has enriched 10 related papers. Consider them for comparative context: