Key result
Non-apical ICD lead position was associated with similar rates of the composite of failed shock or arrhythmic death compared to apical position over 3 years (2.8% vs. 2.3%, P=0.35).
Why the study?
Does non-apical ICD lead position improve or reduce the success of defibrillation compared to apical position in patients undergoing ICD implantation?
Cohort (n=2,475)
Does non-apical ICD lead position improve or reduce the success of defibrillation compared to apical position in patients undergoing ICD implantation?
Non-apical right ventricular lead positioning for ICDs is as effective as apical positioning for both intra-operative defibrillation and long-term clinical shock success.
Supports non-apical RV lead positioning as a viable alternative; leaves open randomized confirmation of long-term equivalence.
INTRODUCTION: We aim to compare the acute and long-term success of defibrillation between non-apical and apical ICD lead position. METHODS AND RESULTS: The position of the ventricular lead was recorded by the implanting physician for 2,475 of 2,500 subjects in the Shockless IMPLant Evaluation (SIMPLE) trial, and subjects were grouped accordingly as non-apical or apical. The success of intra-operative defibrillation testing and of subsequent clinical shocks were compared. Propensity scoring was used to adjust for the impact of differences in baseline variables between these groups. There were 541 leads that were implanted at a non-apical position (21.9%). Patients implanted with a non-apical lead had a higher rate of secondary prevention indication. Non-apical location resulted in a lower mean R-wave amplitude (14.0 vs. 15.2, P < 0.001), lower mean pacing impedance (662 ohm vs. 728 ohm, P < 0.001), and higher mean pacing threshold (0.70 V vs. 0.66 V, P = 0.01). Single-coil leads and cardiac resynchronization devices were used more often in non-apical implants. The success of intra-operative defibrillation was similar between propensity score matched groups (89%). Over a mean follow-up of 3 years, there were no significant differences in the yearly rates of appropriate shock (5.5% vs. 5.4%, P = 0.98), failed appropriate first shock (0.9% vs. 1.0%, P = 0.66), or the composite of failed shock or arrhythmic death (2.8% vs. 2.3% P = 0.35) according to lead location. CONCLUSION: We did not detect any reduction in the ICD efficacy at the time of implant or during follow-up in patients receiving a non-apical RV lead.
No takes yet. Share an insight, caveat, or question.
Amit et al. (2016) conducted a cohort in ICD implantation (n=2,475). Non-apical ICD lead position vs. Apical ICD lead position was evaluated on Composite of failed shock or arrhythmic death (p=0.35). Non-apical ICD lead position was associated with similar rates of the composite of failed shock or arrhythmic death compared to apical position over 3 years (2.8% vs. 2.3%, P=0.35).