Key result
A St. Jude Medical Riata 1570 right ventricular lead exhibited premature insulation failure with inner wire protrusion, causing elevated pacing threshold without impedance change.
Case Report (n=1)
This case highlights a novel mechanism of premature insulation failure in Riata defibrillator leads that can occur without typical warning signs like impedance changes or inappropriate sensing.
Answers are grounded in this paper and cite their sources.
May warrant closer Riata lead surveillance despite normal impedance; leaves open optimal monitoring and intervention strategies.
A 63-year-old woman with a St. Jude Medical Riata 1570 right ventricular lead complained of intermittent hiccups 2 months after implant. Interrogation revealed elevated pacing threshold and diaphragmatic stimulation. Pacing and shock lead impedances remained stable. No inappropriate sensing was noted. Fluoroscopic examination of the lead revealed a thin radio-opaque wire seen between the 2 defibrillator coils away from the main body of the lead. After extraction, a tear in the insulation of the lead was noted allowing the inner wire to protrude. This case illustrates a novel mechanism of insulation failure without inappropriate sensing or impedance change.
No takes yet. Share an insight, caveat, or question.
Krebsbach et al. (2011) conducted a case report in Defibrillator lead failure (n=1). St. Jude Medical Riata 1570 right ventricular lead was evaluated on Insulation failure. A St. Jude Medical Riata 1570 right ventricular lead exhibited premature insulation failure with inner wire protrusion, causing elevated pacing threshold without impedance change.