Key result
A St. Jude Medical Riata 1570 right ventricular lead exhibited premature insulation failure with inner wire protrusion, causing elevated pacing threshold without impedance change.
Population
A 63-year-old woman with a St. Jude Medical Riata 1570 right ventricular lead
Design
Case_report
Follow-up
2 months
Authors
Loading...
May warrant closer Riata lead surveillance despite normal impedance; leaves open optimal monitoring and intervention strategies.
Case Report (n=1)
This case highlights a novel mechanism of premature insulation failure in Riata defibrillator leads that can occur without typical warning signs like impedance changes or inappropriate sensing.
Krebsbach et al. (2011) conducted a case report in Defibrillator lead failure (n=1). St. Jude Medical Riata 1570 right ventricular lead was evaluated on Insulation failure. A St. Jude Medical Riata 1570 right ventricular lead exhibited premature insulation failure with inner wire protrusion, causing elevated pacing threshold without impedance change.