Key result
Defibrillation threshold testing during elective generator change identified an unsuspected short circuit in a Riata 1570 defibrillator lead despite normal interrogation parameters.
Case Report (n=1)
This case highlights the potential utility of routine defibrillator threshold testing during elective generator changes to detect silent lead malfunctions.
Answers are grounded in this paper and cite their sources.
DFT testing at generator change may detect silent Riata lead short circuits; hypothesis-generating and should not yet alter practice.
In this case report we describe a short circuit in the Riata 1570 defibrillator lead (Riata 1570, St. Jude Medical, St. Paul, MN, USA) that was unsuspected owing to normal lead parameters until defibrillator threshold testing at the time of elective generator change. On this occasion, the short circuit manifested as unsuccessful defibrillation of ventricular fibrillation with immediate battery depletion. This report adds weight to existing concerns over narrow caliber leads, it draws attention to the possibility of lead malfunction despite unremarkable interrogation, and lastly it highlights the potential role of routine defibrillator threshold testing, particularly at elective generator change (an issue that remains sparingly addressed in the existing literature).
No takes yet. Share an insight, caveat, or question.
Leong et al. (2012) conducted a case report in Defibrillator lead short circuit (n=1). Defibrillation threshold testing was evaluated on Identification of lead short circuit. Defibrillation threshold testing during elective generator change identified an unsuspected short circuit in a Riata 1570 defibrillator lead despite normal interrogation parameters.
Synapse has enriched 10 related papers. Consider them for comparative context: