Key result
A St. Jude Medical defibrillator with a Riata lead failed to deliver therapy due to internal wire externalization, despite showing normal electrical parameters during routine surveillance.
Case Report (n=1)
This case highlights that St. Jude Medical systems with Riata leads may fail to deliver therapy despite normal electrical parameters during routine surveillance.
Answers are grounded in this paper and cite their sources.
Normal parameters may miss Riata lead failure to deliver therapy; hypothesis-generating for improved integrity surveillance.
We present a case of failure to deliver a shock by a St. Jude Medical defibrillator involving a Riata lead that was discovered incidentally while the device was attempting to deliver inappropriate therapy. Routine interrogation, including high voltage (HV) impedance, failed to reveal any abnormality. Failure to deliver therapy was confirmed during DFT testing, which revealed extremely low HV impedance only while attempting to deliver therapy. Fluoroscopy indicated moderate externalization of internal wires. This case highlights an under-recognized issue with St. Jude Medical systems, namely the possibility that therapy may not be delivered despite the presence of normal electrical parameters during routine surveillance.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2012) conducted a case report in Defibrillator failure (n=1). St. Jude Medical defibrillator with Riata lead was evaluated. A St. Jude Medical defibrillator with a Riata lead failed to deliver therapy due to internal wire externalization, despite showing normal electrical parameters during routine surveillance.