Key result
Routine device interrogation after a defibrillation shock for ventricular fibrillation in a young woman revealed significant lead dysfunction and insulation breaches.
Case Report (n=1)
Thorough evaluation and device interrogation after ICD activations are crucial to unmask potential system failures such as lead insulation breaches.
Routine post-shock ICD interrogation may detect lead insulation breaches; this case report is hypothesis-generating and should not yet change practice.
A young woman with placement of a dual‐chamber implantable cardioverter‐defibrillator (ICD) and a history of prior cardiac arrest due to congenital long QT syndrome presented with defibrillation caused by a ventricular fibrillation arrest. Routine device interrogation revealed significant lead dysfunction. During device revision, breaches were detected in the insulation of both leads within the pre‐pectoral pocket and an “arc mark” was observed on the ICD generator casing; these findings were consistent with a high‐voltage discharge through a short circuit between the denuded right ventricular lead and the casing. In this case, system failure was unmasked only by interrogation after appropriate device activation, which highlights the importance of thorough evaluation after all ICD activations.
No takes yet. Share an insight, caveat, or question.
Eng et al. (2012) conducted a case report in Congenital long QT syndrome with ICD lead dysfunction (n=1). Device interrogation and revision was evaluated on Lead dysfunction and system failure. Routine device interrogation after a defibrillation shock for ventricular fibrillation in a young woman revealed significant lead dysfunction and insulation breaches.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: