Key result
Steroid-eluting pacing leads with 1.2-mm2 electrodes demonstrated a 0% reoperation rate and favorable chronic electrical characteristics over a mean follow-up of 25 months.
Why the study?
Do steroid-eluting pacing leads with 1.2-mm2 electrodes provide favorable electrophysiological parameters and handling characteristics in patients requiring pacemakers?
Observational (n=33)
Do steroid-eluting pacing leads with 1.2-mm2 electrodes provide favorable electrophysiological parameters and handling characteristics in patients requiring pacemakers?
Answers are grounded in this paper and cite their sources.
Steroid-eluting pacing leads with 1.2-mm2 electrodes demonstrate favorable handling, stable low pacing thresholds, and high pacing impedances, potentially increasing pacemaker longevity.
To raise pacing impedance and reduce battery current drain, new tined steroid-eluting leads were developed with 1.2-mm2 hemispherical electrodes, instead of conventional 5-8 mm2. Twenty-two unipolar J-shaped atrial leads and 25 unipolar ventricular leads (models 4533 and 4033, respectively) were implanted in 33 consecutive patients and followed for a mean of 25 months (range 18-29). Handling characteristics of atrial leads were found favorable. The leads slipped easily into the right atrial appendage and were easy to position. Handling characteristics of ventricular leads were satisfying, but more efforts had to be applied to cross the tricuspid valve. Special care was taken to avoid perforation of the myocardium due to the small lead tip. Following implantation, four ventricular and one atrial lead exhibited instability of pacing thresholds that resolved spontaneously within 1-3 days of implantation. Except for this, no lead malfunctioned. The reoperation rate was zero. The mean electrogram amplitudes of 15 mV (ventricle) and 4 mV (atrium), and the mean chronic pacing threshold of 0.085 ms at 1.6 V (app. 0.43 V at 0.5 ms) were comparable with the best values seen in the literature on passive fixation leads. The rest of the electrophysiological parameters were enhanced: mean pacing impedances were 984 omega (acute) and 900 Q (chronic), mean slew rates 3.26 V/s (ventricle) and 1.75 V/s (atrium), mean acute voltage threshold at 0.5 ms was 0.25 V, mean current and energy thresholds calculated at 0.5 ms were 260 microA and 32 nJ (acute) and 478 microA and 103 nJ (chronic). The electrical characteristics of these leads provide for increased pacemaker longevity in combination with substantial safety margins for pacing and sensing.
No takes yet. Share an insight, caveat, or question.
DANILOVIC et al. (1997) conducted an observational in Patients requiring permanent pacing (n=33). Steroid-eluting pacing leads with 1.2-mm2 hemispherical electrodes was evaluated on Reoperation rate. Steroid-eluting pacing leads with 1.2-mm2 electrodes demonstrated a 0% reoperation rate and favorable chronic electrical characteristics over a mean follow-up of 25 months.
Synapse has enriched 10 related papers. Consider them for comparative context: