Key result
Dual chamber ICDs were associated with a lower rate of inappropriate therapy compared to single chamber ICDs.
Why the study?
Does a dual chamber ICD reduce inappropriate ICD therapy compared to a single chamber ICD in patients with an AHA class I indication?
Cohort (n=386)
Does a dual chamber ICD reduce inappropriate ICD therapy compared to a single chamber ICD in patients with an AHA class I indication?
Dual chamber ICDs are associated with a significantly lower rate of inappropriate therapy compared to single chamber ICDs.
Answers are grounded in this paper and cite their sources.
May favor dual-chamber selection to curb inappropriate therapy; leaves open causal benefit in randomized class I cohorts.
INTRODUCTION: Inappropriate implantable cardioverter defibrillator (ICD) therapy (IT) is a common complication in patients with ICD. IT is commonly triggered by supraventricular tachycardias (SVT). Dual chamber ICDs (D-ICDs) may distinguish SVT from ventricular tachycardia/ventricular fibrillation better than single chamber ICDs (S-ICDs) and may be associated with a smaller incidence of IT. METHODS: We reviewed the charts of 386 patients who had an ICD implanted for an AHA class I indication. Intracardiac electrograms were used to classify shocks as either appropriate or inappropriate. RESULTS: Of 295 patients with an S-ICD, 66 (22.3%) received IT, compared to 5 (5.4%) of 91 patients with a D-ICD. The likelihood of being event-free at 1, 2, 3, and 4 years was 96.1%, 96.1%, 96.1%, and 89% for patients with D-ICD and 80.7%, 72.7%, 69.6%, and 66.4%, respectively, for patients with S-ICD (P < 0.001). Multivariate analysis showed no significant association with age, sex, history of atrial fibrillation, history of hypertension, or ejection fraction. SVTs were the commonest cause of IT in our patients. CONCLUSION: Patients with D-ICD are less likely to receive IT as compared to patients with S-ICD.
No takes yet. Share an insight, caveat, or question.
Soundarraj et al. (2006) conducted a cohort in ICD implanted for an AHA class I indication (n=386). Dual chamber ICD (D-ICD) vs. Single chamber ICD (S-ICD) was evaluated on Inappropriate ICD therapy (p=<0.001). Dual chamber ICDs were associated with a lower rate of inappropriate therapy compared to single chamber ICDs.
Synapse has enriched 10 related papers. Consider them for comparative context: