Key result
Automated devices for measuring toe systolic blood pressure demonstrated excellent inter-rater reliability between two podiatrists (ICC 0.85; 95% CI 0.74-0.92).
Why the study?
What is the inter-rater reliability of automated devices for measuring toe systolic blood pressure and toe brachial index?
Observational (n=40)
What is the inter-rater reliability of automated devices for measuring toe systolic blood pressure and toe brachial index?
Automated devices for measuring toe and brachial systolic blood pressure and TBI demonstrate high inter-rater reliability, supporting their use for monitoring lower extremity vascular status.
OBJECTIVES: The aim of this study was to determine the inter-rater reliability of automated devices for the measurement of toe systolic blood pressure (TSBP) and the toe brachial index (TBI). PARTICIPANTS AND METHODS: Toe and brachial systolic blood pressures were measured in 40 participants by two podiatrists using SysToe and MicroLife automated devices. Intraclass correlation coefficients (ICC) with a 95% confidence interval (CI) and 95% limits of agreement (LOA) were determined for TSBP, brachial systolic blood pressure and the TBI. RESULTS: The inter-rater reliability of all measurements was excellent, with ICCs ranging from 0.82 to 0.85. TSBP had the greatest reliability (ICC: 0.85, 95% CI: 0.74-0.92, 95% LOA: -14.4 to 21.1 mmHg) and brachial blood pressures had the lowest reliability (ICC: 0.82, 95% CI: 0.68-0.90, 95% LOA: -21.2 to 22.9 mmHg). The inter-rater reliability of the TBI fell between the TSBP and the brachial blood pressure reliability (ICC: 0.83, 95% CI: 0.70-0.90, 95% LOA: -0.16 to 0.20). CONCLUSION: We showed that automated devices for the measurement of toe and brachial systolic blood pressure and the TBI in a clinical setting produce measurements with high inter-rater reliability. These findings support the use of these automated devices for ongoing monitoring of lower extremity vascular status.
No takes yet. Share an insight, caveat, or question.
Sonter et al. (2014) reported an observational. Automated devices (SysToe and MicroLife) was evaluated on Inter-rater reliability of toe systolic blood pressure (TSBP) (ICC 0.85, 95% CI 0.74-0.92). Automated devices for measuring toe systolic blood pressure demonstrated excellent inter-rater reliability between two podiatrists (ICC 0.85; 95% CI 0.74-0.92).