Key result
Photoplethysmography demonstrates excellent interrater reliability for measuring toe blood pressure in diabetes, reaching ~0.93 ICC.
Why the study?
A reliable tool to measure arterial flow to the feet in people with diabetes is required, and reliability studies for photoplethysmography in this population are lacking.
Is photoplethysmography (PPG) a reliable method for measuring toe blood pressure and toe-brachial index in patients with diabetes mellitus?
Cross-Sectional (n=60)
Single-blind
Randomized limb and first assessor
Yes
Is photoplethysmography (PPG) a reliable method for measuring toe blood pressure and toe-brachial index in patients with diabetes mellitus?
Effect estimate: ICC 0.93 (95% CI 0.89-0.96)
Photoplethysmography is a highly reliable tool for measuring toe blood pressure and toe-brachial index in patients with diabetes, offering a robust alternative to ankle-brachial index.
PPG may offer a reliable toe pressure option in diabetes; leaves open prospective outcome validation before clinical adoption.
BACKGROUND: A reliable tool to measure arterial flow to the feet in people with diabetes is required given that they are particularly prone to peripheral arterial disease. Traditionally, the ankle brachial index (ABI) has been used to measure arterial circulation, but its application is limited due to calcification of larger arteries. More recently, toe pressure and the toe brachial index (TBI) has been suggested as superior to ABI measurements because they assess smaller digital arteries less prone to arterial calcification. However, reliability studies for the clinical use of photoplethysmography (PPG) in people with diabetes are lacking. METHODS: Sixty people with diabetes mellitus (35 males and 25 females, mean age 59.6 yrs) consented to take part in the study. The majority (92%) had type 2 diabetes and 8% had type 1 diabetes. Forty-three percent were diagnosed as having peripheral neuropathy when tested using a biothesiometer and 15% were current smokers (10 - 40/day). A podiatrist and a diabetes educator measured toe and brachial blood pressure independently and in a random order using PPG. These measurements were repeated again seven days later, and subsequently analysed with intraclass correlation coefficients (ICC), 95% confidence intervals (CI) and standard error of measurement (SEM). RESULTS: The intrarater reliability of measuring toe pressures was excellent (ICC3,1 =0.78-0.79, SEM 8 mmHg) and interrater reliability was also excellent (ICC2,2 = 0.93, SEM 4 mmHg). The intrarater reliability for measuring brachial pressures was generally poor (ICC3,1 = 0.40 - 0.42, SEM 19 mmHg) and interrater reliability was fair-good (ICC2,2. 0.65, SEM 14 mmHg). The TBI intrarater reliability was fair-good (ICC3,1 = 0.51-0.72, SEM 0.08), whilst the interrater reliability of TBI was excellent (ICC2,2 = 0.85, SEM 0.07). CONCLUSIONS: Based on these results, interrater and intrarater reliability of PPG is excellent for measuring toe blood pressure, good for TBI and only fair for brachial pressures in people with diabetes mellitus.
No takes yet. Share an insight, caveat, or question.
Scanlon et al. (2012) conducted a cross-sectional in Diabetes mellitus (n=60). Photoplethysmography (Hadeco Smartdop 30EX) was evaluated on Interrater reliability of measuring toe blood pressure (ICC 0.93, 95% CI 0.89-0.96). Photoplethysmography demonstrated excellent interrater reliability (ICC 0.93) and intrarater reliability (ICC 0.78-0.79) for measuring toe blood pressure in adults with diabetes mellitus.