Automated thumb and brachial systolic blood pressures showed only fair correlation, with poor inter-rater reliability and concurrent validity.
Why the study?
This study evaluated the reliability, validity, and comfort of a novel thumb pressure measure and toe-thumb index to determine their suitability as adjuncts or alternatives to ankle-brachial and toe-brachial indices.
Is a novel toe-thumb pressure index a reliable and valid alternative to ankle-brachial and toe-brachial indices for assessing arterial status?
Cross-Sectional (n=34)
Is a novel toe-thumb pressure index a reliable and valid alternative to ankle-brachial and toe-brachial indices for assessing arterial status?
Thumb systolic pressures correlate fairly with brachial pressures and are more comfortable, but wide measurement errors and poor inter-rater reliability limit their current clinical utility.
Answers are grounded in this paper and cite their sources.
Toe-thumb index correlates with toe-brachial but poor reliability limits utility; leaves open need for standardized validation before testing.
AIMS: This study explored the reliability, validity and perceived comfort of a novel thumb pressure measure and calculation of a toe-thumb index to identify their suitability as an adjunct or alternatives to ankle-brachial and toe-brachial indices. METHODS AND RESULTS: Repeated manual thumb and toe systolic blood pressures were conducted using two raters, over two time points, on 34 healthy participants. Concurrent automated toe, thumb and brachial systolic blood pressures as well as comfort ratings for these measures (using a 10 mm visual analogue scale) were captured once by a research assistant. Automated thumb and brachial measures showed fair correlation (ρ = 0.36, p = 0.03) and a toe-thumb index and toe-brachial index good correlation (ρ = 0.62, p < 0.01). Intraclass correlation coefficients (ICC) identified moderate intra-rater reliability for manual thumb pressures for Rater 1 and 2 (ICC 0.57, 95% CI [0.14, 0.79] and ICC 0.74, 95% CI [0.49, 0.87], respectively), while inter-rater reliability was poor (ICC = 0.16, 95% CI [-0.85, 0.47]). Concurrent validity comparing manual and automated measures for thumb pressure was also poor (ICC -0.05, 95% CI [-1.06, 0.72] and ICC 0.42, 95% CI [-0.16, 0.72] Rater 1 and 2 respectively). Thumb measures were significantly more comfortable than brachial measures (5 mm, p < 0.00). CONCLUSION: Thumb systolic pressures are correlated with brachial systolic pressures, with reasonable intra-rater reliability, however, correlation is only fair and measurement error wider than clinically acceptable. Furthermore, manual measures are poorly correlated with automated units. Consequently, caution is required in applying these techniques. As thumb measures were perceived as significantly more comfortable than brachial measures and have an advantage where brachial pressures cannot, or should not, be obtained, further evaluation is warranted.
No takes yet. Share an insight, caveat, or question.
Banwell et al. (2024) conducted a cross-sectional in Healthy (n=34). Thumb pressure measure vs. Brachial pressure measure was evaluated on Correlation between automated thumb and brachial systolic blood pressures (ρ 0.36, p=0.03). Automated thumb and brachial systolic blood pressures showed only fair correlation, with poor inter-rater reliability and concurrent validity.