Key result
PFA preserves electrical integrity in CIED patients, with a 1.3% clinically relevant malfunction rate.
Why the study?
Pulsed field ablation uses high-intensity electric fields that may potentially interact with cardiac implantable electronic devices.
Does pulsed field ablation preserve electrical integrity and avoid device malfunction in patients with cardiac implantable electronic devices?
Cohort (n=157)
Yes
Does pulsed field ablation preserve electrical integrity and avoid device malfunction in patients with cardiac implantable electronic devices?
Pulsed field ablation in patients with CIEDs appears safe with preserved electrical integrity at 3 months, though electromagnetic interference oversensing is common and caution is needed when ablating near device components.
Answers are grounded in this paper and cite their sources.
PFA in CIED patients warrants periprocedural monitoring for EMI and rare malfunctions; leaves open need for larger prospective safety data.
Purpose Pulsed field ablation (PFA) uses high-intensity electric fields to create myocardial lesions and may potentially interact with cardiac implantable electronic devices (CIEDs). This multicenter study evaluated the effects of different PFA technologies on electrical integrity and device-related complications in patients undergoing left- and right-sided atrial ablation. Methods Consecutive patients with CIEDs undergoing catheter ablation with 4 PFA technologies were included. CIED interrogation was performed at baseline, after ablation, and at ≥ 3-month follow-up. Study endpoints were electrical integrity, defined as changes in sensing, pacing threshold, or impedance, and clinically relevant PFA-related CIED malfunction. Real-time device monitoring was performed in a subgroup to assess electromagnetic interference (EMI). Results A total of 157 patients were included. PFA was performed using a pentaspline catheter in 110 patients, a lattice-tip catheter in 26, a variable-loop circular catheter in 16, and a circular array catheter in 5. No significant changes in lead parameters were observed immediately after ablation or at 3-month follow-up, including in patients undergoing right atrial PFA. PFA was frequently associated with EMI-related oversensing (93%). Two patients (1.3%) experienced clinically relevant PFA-related CIED malfunction, consisting of transient reset of a leadless pacemaker and increase of atrial pacing threshold requiring lead revision, both occurring after right atrial lattice-tip PFA. Conclusions PFA for left- and right-sided atrial arrhythmias was associated with preservation of CIED electrical integrity at acute and 3-month follow-up despite frequent EMI-related oversensing. Clinically relevant PFA-related CIED malfunctions were uncommon; however, caution is warranted when delivering PFA in close proximity to CIED components. Capsule summary In this multicenter study, PFA of left- and right-sided atrial arrhythmias using four contemporary technologies demonstrated preserved CIED electrical integrity at acute and 3-month follow-up despite frequent EMI-related oversensing. Clinically relevant PFA-related CIED malfunctions were uncommon (1.3%), although caution is warranted when delivering PFA in close proximity to CIED components.

J Interv Card Electrophysiol · cohort. Device patients undergoing PFA: EMI oversensing in 93% of monitored cases. Pulsed field ablation of atrial arrhythmias in patients with cardiac implantable electronic devices: a multicenter evaluation of electrical integrity and device safety 157 CIED patients: Pulsed field ablation for atrial arrhythmias. Lead parameters unchanged acutely and at 3 months: Sensing, Pacing threshold, Impedance. 1.3% of patients: Clinically relevant CIED malfunction in 2 patients. Both after right atrial lattice-tip PFA: Leadless pacemaker reset, Atrial threshold increase. Bottom line: PFA preserved device integrity; monitor for EMI and rare malfunctions.
No takes yet. Share an insight, caveat, or question.
Monte et al. (2026) conducted a cohort in Atrial arrhythmias in patients with cardiac implantable electronic devices (n=157). Pulsed field ablation (PFA) was evaluated on Electrical integrity (changes in sensing, pacing threshold, or impedance) and clinically relevant PFA-related CIED malfunction. Pulsed field ablation for atrial arrhythmias in patients with cardiac implantable electronic devices preserved electrical integrity, with clinically relevant CIED malfunctions occurring in 1.3%.