Why the study?
Pulsed field ablation uses high-intensity electric fields that may potentially interact with cardiac implantable electronic devices.
Does pulsed field ablation preserve electrical integrity and avoid device malfunction in patients with cardiac implantable electronic devices?
Population
157 consecutive patients with CIEDs undergoing atrial catheter ablation
Comparison
Catheter ablation using 4 different PFA technologies
Design
Multicenter study
Follow-up
≥ 3-month
Key result
Pulsed field ablation for atrial arrhythmias in patients with cardiac implantable electronic devices preserved electrical integrity, with clinically relevant CIED malfunctions occurring in 1.3%.
Authors
Loading...
PFA in CIED patients warrants periprocedural monitoring for EMI and rare malfunctions; leaves open need for larger prospective safety data.
Cohort (n=157)
Yes
Does pulsed field ablation preserve electrical integrity and avoid device malfunction in patients with cardiac implantable electronic devices?
Pulsed field ablation in patients with CIEDs appears safe with preserved electrical integrity at 3 months, though electromagnetic interference oversensing is common and caution is needed when ablating near device components.
Monte et al. (2026) conducted a cohort in Atrial arrhythmias in patients with cardiac implantable electronic devices (n=157). Pulsed field ablation (PFA) was evaluated on Electrical integrity (changes in sensing, pacing threshold, or impedance) and clinically relevant PFA-related CIED malfunction. Pulsed field ablation for atrial arrhythmias in patients with cardiac implantable electronic devices preserved electrical integrity, with clinically relevant CIED malfunctions occurring in 1.3%.