Key result
Increasing the lower pacemaker rate to 60 bpm successfully overrode the underlying junctional rhythm and resolved the inappropriate ventricular safety pacing.
Case Report (n=1)
Inappropriate ventricular safety pacing triggered by an underlying junctional rhythm can be resolved by simply increasing the lower pacemaker rate.
Answers are grounded in this paper and cite their sources.
Simple rate increase may resolve junctional rhythm-triggered safety pacing; hypothesis-generating and requires prospective validation.
Ventricular safety pacing (VSP) is used to avoid cross talk by delivering ventricular stimulus shortly after an atrial-paced event if ventricular-sensed event occurs. Although VSP is a protective feature that exists for decades in different pacing devices, there are some reports of unfavorable outcomes of this algorithm. More so, health care providers sometimes face difficulties in interpreting and dealing with VSP strips. This case report discusses an important pacemaker algorithm and encourages further attention to possible pitfalls and hence avoids unnecessary interventions.
No takes yet. Share an insight, caveat, or question.
Shurrab et al. (2017) conducted a case report in Ventricular safety pacing anomaly (n=1). Increasing lower pacemaker rate vs. 50 bpm was evaluated. Increasing the lower pacemaker rate to 60 bpm successfully overrode the underlying junctional rhythm and resolved the inappropriate ventricular safety pacing.