Key result
A micropower motion artifact estimator fabricated in a 0.18-μm CMOS process occupies an active area of 0.11 mm2 and consumes 2.6-3.2 μW of power.
Why the study?
Compact low-power motion artifact estimation was needed to reduce input dynamic range in wearable ECG acquisition systems.
The proposed micropower motion artifact estimator offers a low-power, small-area solution for reducing input dynamic range in wearable ECG systems.
Answers are grounded in this paper and cite their sources.
May advance motion-artifact mitigation in ambulatory ECG; leaves open prospective clinical validation.
This work presents the design and analysis of a compact low-power motion artifact estimator for reducing the input dynamic range in wearable ECG acquisition systems. The estimator employs a novel mixed-signal architecture that performs adaptive filtering on the electrode impedance information to derive a cancellation signal to suppress the motion artifact at the input of the acquisition system. A detailed noise analysis and optimization strategies are also provided to minimize the estimator's noise referred to the acquisition system's input. Fabricated in a 0.18- μm CMOS process and operating from a 1-V supply, the estimator occupies an active area of 0.11 [Formula: see text] and consumes 2.6-3.2 μW of power depending on the final weights used. The low power consumption and small area thus make the estimator suitable for local placement at each recording channel in multi-channel ECG acquisition systems.
No takes yet. Share an insight, caveat, or question.
Pholpoke et al. (2019) studied this question. Micropower motion artifact estimator was evaluated on Power consumption and active area. A micropower motion artifact estimator fabricated in a 0.18-μm CMOS process occupies an active area of 0.11 mm2 and consumes 2.6-3.2 μW of power.
Synapse has enriched 10 related papers. Consider them for comparative context: