Key result
Changes in the QRS-complex and skeletal muscle potentials are more important causes of pacemaker failure than external interference, while electronic component defects may occur in 1 of 1000 generators.
Why the study?
What are the primary factors affecting cardiac pacemaker function and causing pacemaker failure?
What are the primary factors affecting cardiac pacemaker function and causing pacemaker failure?
Answers are grounded in this paper and cite their sources.
This review highlights that internal physiological factors and manufacturing defects are more common causes of pacemaker failure than external interference.
A survey is given of the factors affecting cardiac pacemaker function. Whereas it was earlier considered that external interference was a frequent cause of a pacemaker failure, more recent studies indicate that this is of minor importance. It would appear that failure of pacemaker function due to changes in the QRS-complex, i.e. voltage change, intraventricular conduction defects and frequency changes, intraventricular conduction defects and frequency changes, are more important. A lesser known cause of failure in demand function is the inhibition which arises from skeletal muscle postentials. This appears to be a major problem in some types of pacemaker. Manufacturing defects are unavoidable, and it can still be assumed that electronic component defects may develop in 1 of 1000 pulse generators. Cases with the problem of early run-away pacemaker are still being reported and two such cases are discussed. In one of the patients the run-away phenomenon was intermittent, and was accompanied by a variation in run-away frequency. This possibility should thus always be taken into consideration in a pacemaker patient presenting with snycope. In a patient with a QRS-inhibited pacemaker a double stimulation phenomanen has been observed.
No takes yet. Share an insight, caveat, or question.
Ole‐Jørgen Ohm (1976) conducted a review in Cardiac pacemaker failure. Factors affecting cardiac pacemaker function was evaluated. Changes in the QRS-complex and skeletal muscle potentials are more important causes of pacemaker failure than external interference, while electronic component defects may occur in 1 of 1000 generators.