Key result
A new automatic threshold tracking algorithm correctly verified capture or loss of capture for every single analyzed beat across different pacing outputs in all 35 patients.
Why the study?
Does a new automatic threshold tracking algorithm correctly verify ventricular capture in patients with unipolar pacemaker leads?
Population
35 patients requiring pacemakers (26 with new implants and 9 with chronic leads)
Comparison
A new automatic threshold tracking algorithm… vs Tape-recorded intracardiac heart signals
Design
Cohort
Authors
Loading...
Extends automatic threshold tracking validation to unipolar leads; hypothesis-generating pending larger prospective studies.
Does a new automatic threshold tracking algorithm correctly verify ventricular capture in patients with unipolar pacemaker leads?
A novel automatic threshold tracking algorithm accurately verifies ventricular capture beat-by-beat using unipolar leads, potentially expanding the use of automatic threshold tracking beyond low polarization bipolar leads.
Kennergren et al. (2003) studied Pacemaker implantation (n=35). New automatic threshold tracking algorithm for unipolar leads was evaluated on Verification of capture or loss of capture. A new automatic threshold tracking algorithm correctly verified capture or loss of capture for every single analyzed beat across different pacing outputs in all 35 patients.