Key result
The ventricular automatic capture feature accurately determined stimulation thresholds compared to manual testing across 31 different lead types.
Why the study?
Does the ventricular automatic capture feature accurately determine the ventricular stimulation threshold and safely deliver backup pulses in patients with pacemakers?
Observational (n=105)
Yes
Does the ventricular automatic capture feature accurately determine the ventricular stimulation threshold and safely deliver backup pulses in patients with pacemakers?
Answers are grounded in this paper and cite their sources.
The ventricular automatic capture feature accurately determines stimulation thresholds and safely delivers backup pulses across a variety of ventricular leads.
BACKGROUND: This study was conducted to evaluate the clinical performance of the ventricular automatic capture feature as implemented in the Insignia I Ultra pacemaker system (Guidant) utilizing a variety of ventricular leads. Currently, the optimal programming of the pacemaker output considers both pacemaker efficiency (prolonging battery longevity) and patient safety (adequate safety margin). The ability of a pacemaker to automatically adjust the ventricular output above the pacing threshold while maintaining the appropriate safety margin has been explored since the early 1970s and is only available today in conjunction with a specific low polarization lead system. METHODS: One hundred and five patients were enrolled from 17 European centers utilizing 31 different types of ventricular leads were followed through their 3-month follow-up visit. There were no restrictions on the type of ventricular leads used. RESULTS: The average mean difference between the commanded autothreshold test (0.652 + 0.335 V) and the manual threshold test (0.651 + 0.335 V) was 0.001 + 0.49 (P < 0.0001). The average mean difference between the ambulatory autothreshold test (0.696 + 0.322 V) and the commanded autothreshold test (0.682 + 0.315 V) was 0.002 + 0.74 (P < 0.0001). Holter recordings confirmed that there were no loss of capture incidences without a backup pulse being delivered. In addition, the mean number of backup pulses delivered in a 24-hour period was less than 0.1% of the total number of paced beats. CONCLUSIONS: This study provided that the automatic capture feature while using a variety of leads accurately determines the ventricular stimulation threshold and safely delivers a backup pulse when required.
No takes yet. Share an insight, caveat, or question.
Sperzel et al. (2005) conducted an observational in Pacemaker indication (n=105). Ventricular automatic capture feature (Insignia I Ultra pacemaker) vs. Manual threshold test was evaluated on Difference between commanded autothreshold test and manual threshold test (MD 0.001, p=< 0.0001). The ventricular automatic capture feature accurately determined stimulation thresholds compared to manual testing across 31 different lead types.