Key result
The automatic capture algorithm significantly increased projected device longevity compared to standard programming (10.3 vs 8.9 years; P<0.0001).
Why the study?
Does an automatic capture algorithm improve projected device longevity in patients implanted with a pacemaker?
RCT (n=910)
1:1
Does an automatic capture algorithm improve projected device longevity in patients implanted with a pacemaker?
Answers are grounded in this paper and cite their sources.
Automatic capture algorithms in pacemakers significantly increase projected device longevity by reducing average ventricular voltage output compared to standard programming.
Background: An automatic capture (AC) algorithm adjusts ventricular pacing output to capture the ventricle while optimizing output to 0.5 V above threshold. AC maintains this output and confirms capture on a beat‐to‐beat basis in bipolar and unipolar pacing and sensing. Objective: To assess the AC algorithm and its impact on device longevity. Methods: Patients implanted with a pacemaker were randomized 1:1 to have the AC feature on or off for 12 months. Two threshold tests were conducted at each visit— automatic threshold and manual threshold. Average ventricular voltage output and projected device longevity were compared between AC on and off using nonparametric tests. Results: Nine hundred ten patients were enrolled and underwent device implantation. Average ventricular voltage output was 1.6 V for the AC on arm (n = 444) and 3.1 V for the AC off arm (n = 446) (P < 0.001). Projected device longevity was 10.3 years for AC on and 8.9 years for AC off (P < 0.0001), or a 16% increase in longevity for AC on. The proportion of patients in whom there was a difference between automatic threshold and manual threshold of ≤0.5 V through follow‐up was 99.0% (95% CI: 98.6, 99.3). The average difference between automatic threshold and manual threshold was 0.07 V (P = 0.002). Conclusions: This study showed that automatic threshold testing at follow‐up provided comparable information to manually measured threshold testing. AC provided a significant increase in projected device longevity compared to standard programing.
No takes yet. Share an insight, caveat, or question.
Koplan et al. (2008) conducted an RCT in Pacemaker implantation (n=910). Automatic capture (AC) algorithm vs. AC off (standard programming) was evaluated on Projected device longevity (16% increase, p=<0.0001). The automatic capture algorithm significantly increased projected device longevity compared to standard programming (10.3 vs 8.9 years; P<0.0001).