Key result
AutoCapture extended estimated pacemaker longevity compared with nominal settings, especially in patients with high pacing thresholds.
Why the study?
Does the AutoCapture algorithm improve pacemaker longevity compared to nominal or optimized manual output in patients with St. Jude Affinity series pacemakers?
Does the AutoCapture algorithm improve pacemaker longevity compared to nominal or optimized manual output in patients with St. Jude Affinity series pacemakers?
The AutoCapture algorithm significantly prolongs pacemaker battery longevity compared to nominal settings, particularly in patients with high pacing thresholds.
Answers are grounded in this paper and cite their sources.
May support selective AutoCapture use to extend device life in high-threshold patients; leaves open randomized confirmation of longevity gains.
BACKGROUND: The automatic, threshold tracking, pacing algorithm developed by St. Jude Medical, verifies ventricular capture beat by beat by recognizing the evoked response following each pacemaker stimulus. This function was assumed to be not only energy saving but safe. This study estimated the extension in longevity obtained by AutoCapture (AC) compared with pacemakers programmed to manually optimized, nominal output. METHODS: Thirty-four patients who received the St. Jude Affinity series pacemaker were included in the study. The following measurements were taken: stimulation and sensing threshold, impedance of leads, evoked response and polarization signals by 3501 programmer during followup, battery current and battery impedance under different conditions. For longevity comparison, ventricular output was programmed under three different conditions: (1) AC on; (2) AC off with nominal output, and (3) AC off with pacing output set at twice the pacing threshold with a minimum of 2.0 V. Patients were divided into two groups: chronic threshold is higher or lower than 1 V. The efficacy of AC was evaluated. RESULTS: Current drain in the AC on group, AC off with optimized programming or nominal output was (14.33 +/- 2.84) mA, (16.74 +/- 2.75) mA and (18.4 +/- 2.44) mA, respectively (AC on or AC off with optimized programming vs. nominal output, P < 0.01). Estimated longevity was significantly extended by AC on when compared with nominal setting [(103 +/- 27) months, (80 +/- 24) months, P < 0.01). Furthermore, compared with the optimized programming, AC extends the longevity when the pacing threshold is higher than 1 V. CONCLUSION: AC could significantly prolong pacemaker longevity; especially in the patient with high pacing threshold.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2006) studied Pacemaker implantation (n=34). AutoCapture (AC) on vs. AC off with nominal output was evaluated on Estimated longevity in months (p=< 0.01). AutoCapture extended estimated pacemaker longevity compared with nominal settings, especially in patients with high pacing thresholds.