Key result
AutoCapture significantly extended estimated pacemaker longevity compared with nominal settings (103 vs 80 months, P<0.01), especially in patients with high pacing thresholds.
Why the study?
Does the AutoCapture algorithm improve pacemaker longevity compared to nominal or optimized manual output in patients with St. Jude Affinity series pacemakers?
Population
34 patients who received the St. Jude Affinity series pacemaker
Comparison
AutoCapture (AC) algorithm turned on vs AutoCapture off with nominal output, and…
Design
Cohort
Authors
Loading...
May support selective AutoCapture use to extend device life in high-threshold patients; leaves open randomized confirmation of longevity gains.
Does the AutoCapture algorithm improve pacemaker longevity compared to nominal or optimized manual output in patients with St. Jude Affinity series pacemakers?
Absolute Event Rate: 103% vs 80%
p-value: p=< 0.01
The AutoCapture algorithm significantly prolongs pacemaker battery longevity compared to nominal settings, particularly in patients with high pacing thresholds.
Chen et al. (2006) studied Pacemaker implantation (n=34). AutoCapture (AC) on vs. AC off with nominal output was evaluated on Estimated longevity in months (p=< 0.01). AutoCapture significantly extended estimated pacemaker longevity compared with nominal settings (103 vs 80 months, P<0.01), especially in patients with high pacing thresholds.