Key result
The Autocapture function was projected to increase pacemaker longevity by 53% to 245% and reduce pacing-related costs by 25% to 57% over 15 years compared to standard settings.
Why the study?
Does automatic threshold tracking improve projected pacemaker longevity and reduce pacing related costs in patients with VVIR pacemakers?
Observational (n=25)
Does automatic threshold tracking improve projected pacemaker longevity and reduce pacing related costs in patients with VVIR pacemakers?
Automatic threshold tracking in pacemakers significantly increases projected battery longevity and reduces pacing-related costs compared to standard fixed-output settings.
Answers are grounded in this paper and cite their sources.
Projected longevity and cost benefits with Autocapture should not yet alter VVIR practice; leaves open need for randomized confirmation.
The aim of this study was to calculate the battery current drain of different VVI-VVIR pulse generators, with and without automatic adjustment of stimulation output, from the same manufacturer, and obtain projections on pacemaker longevity and pacing related costs. Twenty-five patients who had received a VVIR pacemaker (Microny SR+ 2425T) with Autocapture function, an algorithm able to provide automatic adaptation of device output coupled with continuously monitored capture threshold, were analyzed. Calculations of current drain and projections of device longevity were based on diagnostic data retrieved from device memory and on comparisons of multiple programmed settings of the same device (Microny SR+ 2425T with Autocapture, Low Output [output = 2 x Autocapture threshold, but not less than 2.4 V] versus Standard Output [5 V]). In addition, devices with Autocapture but with different battery capacities (0.35 Ah and 0.79 Ah, Microny and Regency, respectively) were compared to a standard device (battery capacity = 0.78 Ah) at 2.5-V and 5-V output settings. According to a series of assumptions, Autocapture increased device longevity by 53% (Microny) and by 245% (Regency) over a 15-year follow-up with a consequent reduction in pacing related costs of 25% (Microny) and of 57% (Regency) compared to the standard settings of a VVIR device from the same manufacturer. In conclusion, pacing with constant adaptation of generator output coupled with continuous monitoring of stimulation threshold has the potential to increase generator longevity and to reduce pacing related costs as compared to a standard device from the same manufacturer.
No takes yet. Share an insight, caveat, or question.
Boriani et al. (2000) conducted an observational in Patients requiring a VVIR pacemaker (n=25). Autocapture function (automatic adaptation of stimulation output) vs. Standard Output (5 V) settings / standard device was evaluated on Projected pacemaker longevity and pacing related costs (53% to 245% increase in longevity; 25% to 57% reduction in costs). The Autocapture function was projected to increase pacemaker longevity by 53% to 245% and reduce pacing-related costs by 25% to 57% over 15 years compared to standard settings.