Key result
The ACAP Confirm algorithm successfully managed atrial capture in 65% of newly implanted patients pre-discharge, which significantly improved to 88% during medium-term follow-up (p<0.001).
Why the study?
Does the ACAP Confirm algorithm accurately and reliably measure atrial capture thresholds compared to manual testing in patients with pacemakers or ICDs?
Observational (n=318)
Yes
Does the ACAP Confirm algorithm accurately and reliably measure atrial capture thresholds compared to manual testing in patients with pacemakers or ICDs?
Absolute Event Rate: 88% vs 65%
p-value: p=<0.001
The ACAP Confirm algorithm provides reliable automatic atrial capture threshold measurements that correlate well with manual testing, with reliability increasing over the medium term after implantation.
May support automated atrial capture management after implantation; leaves open randomized validation versus manual testing.
Background Device‐based fully automatic pacing capture detection is useful in clinical practice and important in the era of remote care management. The main objective of this study was to verify the effectiveness of the new ACAP Confirm® algorithm in managing atrial capture in the medium term in comparison with early post‐implantation testing. Methods Data were collected from 318 patients (66% male; mean age, 73±10 years); 237 of these patients underwent device implantation and 81 box changes in 31 Italian hospitals. Atrial threshold measurements were taken manually and automatically at different pulse widths before discharge and during follow‐up (7±2 months) examination. Results The algorithm worked as expected in 73% of cases, considering all performed tests. The success rate was 65% and 88% pre‐discharge and during follow‐up examination (p<0.001), respectively, in patients who had undergone implantation. We did not detect any difference in the performance of the algorithm as a result of the type of atrial lead used. The success rate was 70% during pre‐discharge testing in patients undergoing device replacement. Considering all examination types, manual and automatic measurements yielded threshold values of 1.07±0.47 V and 1.03±0.47 V at 0.2‐ms pulse duration (p=0.37); 0.66±0.37 V and 0.67±0.36 V at 0.4 ms (p=0.42); and 0.5±0.28 V and 0.5±0.29 V at 1 ms (p=0.32). Conclusions The results show that the algorithm works before discharge, and its reliability increases over the medium term. The algorithm also proved accurate in detecting the atrial threshold automatically. The possibility of activating it does not seem to be influenced by the lead type used, but by the time from implantation.
No takes yet. Share an insight, caveat, or question.
Quirino et al. (2016) conducted an observational in Patients requiring permanent pacing with atrial leads (n=318). ACAP Confirm algorithm vs. Pre-discharge testing (baseline) was evaluated on Successful activation of the ACAP Confirm algorithm by means of the in-clinic automatic threshold test (p=<0.001). The ACAP Confirm algorithm successfully managed atrial capture in 65% of newly implanted patients pre-discharge, which significantly improved to 88% during medium-term follow-up (p<0.001).