Key result
Automatic atrial threshold monitoring strictly correlates with manual physician assessments over 8 months.
Why the study?
Does the ambulatory automatic atrial threshold monitoring algorithm accurately measure atrial pacing thresholds compared to manual physician assessment in patients with pacemakers?
Observational (n=352)
Yes
Does the ambulatory automatic atrial threshold monitoring algorithm accurately measure atrial pacing thresholds compared to manual physician assessment in patients with pacemakers?
p-value: p=<0.001
The automatic atrial threshold monitoring algorithm in pacemakers is highly feasible and correlates excellently with manual physician measurements, supporting its use in routine clinical practice.
Supports automatic atrial threshold monitoring in pacemaker patients; leaves open randomized confirmation before routine adoption.
AIMS: The aim of this study was to validate the ambulatory automatic atrial threshold monitoring algorithm by comparing the measurements assessed by the automatic system and those evaluated manually by the physician at discharge, 2- and 8-month follow-up sessions. METHODS AND RESULTS: This is an observational multicentric prospective study of 352 patients implanted with EnPulse(®) DR pacemakers. Mean age was 76.3 ± 9.4 years. Indications of pacing were atrio-ventricular block (AVB) (64%) and sinus dysfunction (SD) or brady-tachy syndrome (36%). The automatic atrial threshold monitoring function was maintained at nominal programming state with daily measurement scheduled at 1:00 am. Ambulatory automatic atrial threshold assessment was possible for 91.5% of patients at discharge, 97.3% at 2 months, and 95.7% at 8 months. Causes of the unsuccessful attempts to perform automatic atrial threshold were atrial arrhythmias or permanent atrial and ventricular pacing. Feasibility is significantly better for AVB indication than SD indication due to more frequent occurrence of atrial fibrillation (AF). At each stage, there is a strict correlation between the automatic measurements and those conducted manually by the physician with a P < 0.001. CONCLUSION: Feasibility of ambulatory automatic atrial threshold is good. Results of the study show excellent correlation between the two methods for atrial threshold: there is no statistical difference between manual and automatic measurements during follow-up.
No takes yet. Share an insight, caveat, or question.
Rey et al. (2012) conducted an observational in Patients implanted with EnPulse DR pacemakers (n=352). Ambulatory automatic atrial threshold monitoring algorithm vs. Manual measurement by physician was evaluated on Correlation between automatic and manual measurements of atrial pacing threshold (p=<0.001). Ambulatory automatic atrial threshold measurements showed a strict correlation with manual physician assessments during 8 months of follow-up (P<0.001).