Key result
Atrial automatic capture verification confirms ~100% of paced beats using independent electrodes.
Why the study?
Does an atrial automatic capture verification scheme using independent pacing and sensing electrodes accurately detect atrial evoked response in pacemaker patients?
Does an atrial automatic capture verification scheme using independent pacing and sensing electrodes accurately detect atrial evoked response in pacemaker patients?
Atrial automatic capture verification by atrial evoked response detection is highly accurate and feasible using independent pacing and sensing electrodes.
Answers are grounded in this paper and cite their sources.
Supports atrial capture verification feasibility; leaves open outcome benefits in prospective trials.
INTRODUCTION: This study evaluated an atrial automatic capture verification scheme based on atrial evoked response (AER). Atrial pacing was between Atip and Can (Atip-Can) using different coupling capacitances (CCs). Independent pairs of sensing electrodes between Aring and Vtip (Aring-Vtip) or between Aring and a separate indifferent electrode (Aring-Indiff) were used to reduce pacing-induced afterpotentials. METHODS AND RESULTS: A custom-made external pacing system was used to perform automatic step-up and step-down pacing (0.1 to 7.1 V at 0.5 msec, step size of 0.1 V) using different CCs (2 or 15 microF). Intracardiac signals from Aring-Indiff and Aring-Vtip were independently recorded and analyzed both in real time and off-line to detect AER. Every paced beat also was visually inspected and compared with surface ECG to verify the captures. With the intracardiac signals properly filtered, AER detection was based on the signal within a window of 12 to 65 msec after the stimulus. Data from 27 patients (4 chronic and 23 acute implantations; age 65.6+/-13.9 years) were analyzed. Bipolar atrial lead measurements using a standard pacing system analyzer were as follows (mean +/- SD): impedance 695+/-227 ohms, P wave amplitude 4.2+/-2.3 mV, slew rate 1.1+/-0.9 V/sec, and pacing threshold at 0.5 msec 1.0+/-0.5 V. The results with CC = 2 microF showed that of 9,500 atrial paced beats, correct capture verification rates were 99.8% (Aring-Indiff) and 99.4% (Aring-Vtip). Similar results were achieved with CC = 15 microF (99.7% and 99.5%, respectively). CONCLUSION: AER can be reliably detected using independent pacing (Atip-Can) and sensing (Aring-Vtip or Aring-Indiff) electrodes. Therefore, atrial automatic capture verification by AER detection is feasible.
No takes yet. Share an insight, caveat, or question.
Butter et al. (2001) studied Pacemaker implantation (n=27). Atrial automatic capture verification based on atrial evoked response vs. Visual inspection and surface ECG was evaluated on Correct capture verification rate. Atrial automatic capture verification based on atrial evoked response correctly verified capture in 99.4% to 99.8% of paced beats using independent pacing and sensing electrodes.