Does an external pacing system with reduced coupling capacitance allow adequate evoked-response detection for automatic capture verification in patients with ICD lead systems?
Does an external pacing system with reduced coupling capacitance allow adequate evoked-response detection for automatic capture verification in patients with ICD lead systems?
Reducing the pacing coupling capacitance allows adequate evoked-response detection for automatic capture verification on non-selected ICD lead systems.
Answers are grounded in this paper and cite their sources.
Successful evoked-response detection achieved with reduced-capacitance pacing across ICD leads; extends feasibility data but remains hypothesis-generating.
AIM: Intracardiac ventricular evoked-response (ER) signals detected by implanted cardioverter defibrillator (ICD) lead systems were investigated for automatic capture verification (AC). METHOD: ER signals were evaluated with an external pacing system equipped with a reduced coupling capacitance (CC=2.2 microF) in the pacing output circuit during ventricular step-down threshold testing at 0.4 ms pulse duration. Real-time pacing markers, surface ECG and intracardiac electrograms pre- and post-filtering were digitally recorded. RESULTS: Twenty consecutive patients, age 61+/-12 years, with leads from two different manufacturers were tested - 10 were implanted with acute leads (AL) and 10 with chronic leads (CL). The analysis was based on the ER amplitude during capture and on the ER-to-afterpotential ratio (SAR), with SAR>2 as the criterion for successful capture detection. ER amplitudes (median and range) were 8.1 mV (2.1-19.5 mV) for AL and 8.3 mV (3.7-14.2 mV) for CL. SAR values (median and range) were 48.0 (2.5-682.6) for AL and 13.2 (6.3-35.9) for CL, indicating that AC could successfully be applied in all patients. CONCLUSIONS: Reducing the pacing CC allows adequate ER detection for automatic capture verification on non-selected ICD lead systems. The effect of high-voltage shock treatment on ER signal detection requires further investigation.
No takes yet. Share an insight, caveat, or question.
Johannes Sperzel (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: